Selecting stop bits:
Full duplex/half duplex:
The processor does not allow the following combinations:
no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP230 program will change the number of stop bits to 2
even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the program changes the number of stop bits to 1
NOTE
The serial bus settings become effective only after reset.
6.3.5.3. Setting the transmitter address
Address is used when more than one transmitter is connected to one serial bus;
this way, it is possible to communicate with one transmitter at a time.
Select Addr in the main menu; the following is displayed:
Pressing ENT returns the program to the main menu.
Pressing CL deletes the old address; enter the new address with the
arrow keys.