6.4.2. Output via the serial bus
6.4.2.1. Starting the measurement output
R <cr>
Starts output of measurements to the peripheral devices (RUN mode); the only
command that can be used is S (stop).
The output mode can be changed with command FORM (see Appendix 1).
6.4.2.2. Stopping the measurement output
S<cr>
Ends the RUN mode; after this command all other commands can be used.
6.4.2.3. Outputting the reading once
SEND <cr> in STOP mode
or
SEND aa <cr> in POLL state
aa = address of the transmitter when more than one transmitter
is connected to a serial bus (0...99)
The output format depends on which parameters the transmitter can output.
Output types:
RH= 21.9 %RH T= 23.9 'C
RH= 21.9 %RH T= 23.9 'C Td=0.9 'C
RH= 21.9 %RH T=23.9 'C a=4.7 g/m3 x=4.0 g/kg Tw= 12.3 'C
RH= 21.9 %RH T= 23.9 'C Td=0.9 'C a=4.8 g/m3 x=4.0 g/kg
Tw=12.3 'C h=34.4 kJ/kg
RH= 21.9 %RH T= 24.0 'C h= 34.4 kJ/kg
The output mode can be changed with command FORM (see Appendix 1).