Instruction Manual

HMP230 SERIES
User's Guide M210225en-B
58
Give command FCRH <cr>, enter the first point value and press <cr>:
>FCRH <cr>
RH : xx.x Ref1 ? yy.y <cr>
Press any key when ready...
The stabilization of the sensor to the reference humidity can be
monitored by giving c <cr>:
RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...
Insert the sensor head into a measurement hole of the NaCl chamber.
Wait at least 30 minutes.
Press any key and enter the second point value and press <cr>.
RH : xx.x Ref2 ? yy.y <cr>
The stabilization of the sensor can be monitored by entering c <cr>
instead of the reference value.
7.1.3.2. With display/keypad commands
Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.
Disconnect the security lock jumper!
Insert the sensor head into a measurement hole of the LiCl chamber.
Wait at least 30 minutes.
Select Cali in the main menu and then RH cal; select Sensor
changed. Change the first point reading with the arrow keys and
press ENT.
Insert the sensor head into a measurement hole of the NaCl chamber.
Wait at least 30 minutes.
If necessary, change the second point reading with the arrow keys and
press ENT.
7.1.3.3. With LED commands
Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.
Disconnect the security lock jumper!
Insert the sensor head into a measurement hole of the LiCl chamber.