Specifications

100 Commerce Way, Woburn, MA 01801
TEL: 1-888-VAISALA (824-7252) FAX: (781) 933-8029
E-MAIL: incsales@vaisala.com Access catalog on-line at: www.vaisala.com/inc/ssdcat
54
TECHNICAL DATA - HUMITTER
Relative Humidity
Measuring range (for which
accuracy is specified) 10...90% RH
Accuracy at 68°F (+20°C) better than ±3% RH
Operating range 0...100% RH
Stability: ±2% RH over 2 years
Temperature dependence < ±1.5% RH from -10° to +60°C
Sensor INTERCAP
®
part no. 15778
Temperature
Y model -active output
Measuring range +14...140 °F (-10...+60°C)
Accuracy at 68°F (+20°C) 1.1°F (± 0.6 °C)
Sensor PT 1000 IEC 751 Class B
YX model - passive output
2-wire connection wire 2 x 0.08 ohm
resistance
Sensor PT 1000 IEC 751 Class B
General
Input power 7...28 VDC
Output signal 0...1 V, R
L
> 100 ký,
(equals 0...100% RH and
-40...140 °F (-40...+60°C) (10mV equals
one % RH or °C)
0...5 V available for model
50 U only.*
other voltage outputs and
4...20 mA output available
when using external
electronics
YX Model has only passive resistive output corresponding to
PT 1000 specification
Current consumption 2 mA typical
Operating temperature range +14...+140°F (-10...+60°C)
Storage temperature range -40...140 °F (-40...+60°C)
Operating humidity range 0 ... 100% RH
Sensor protection membrane filter,
part no. 16131
option plastic grid, part no. 15724
Housing material and
classification ABS plastic, NEMA 4 (IP 65)
Dimensions ø 12 mm, length 69 mm
Cable length 313 mm
Power supply 7...28 VDC
* Input power: 11-16 VDC
Specifications subject to change without notice.
INTERCAP
®
is a registered trademark of VAISALA, Inc.
HUMITTER
®
Series
Dimensions in mm
18
Ø 11.8
69
313
51
10
M12X1
Ø 12
In HUMITTER 50YX resistive temperature
output via the violet and white wires
Cable connections
BACK to
Table of
Contents
To INDEX
HUMITTER
Price List