Specifications

100 Commerce Way, Woburn, MA 01801
TEL: 1-888-VAISALA (824-7252) FAX: (781) 933-8029
E-MAIL: incsales@vaisala.com Access catalog on-line at: www.vaisala.com/inc/ssdcat
62
HMR 1900 CHART RECORDER
Recording 1,2,3,or 4 pens (red, blue, green, black)
3 position real-time event pen
10 dia or 105mm radius charts
1 hour to 32 days per revolution
Analog Inputs
signal types: mV, V, mA, ohm
Pt 100 resistance thermometer
Accuracy ±0.1% of reading
Transmitter power supply
1 per channel
24 Vdc
Analog outputs 4...20 mA
±0.1% accuracy
Relay outputs
SPDT
Rating 5A at 115/230 Vac
Operating range 32° to 130 °F (0° to 55°C)
5 to 95 %RH (non-condensing)
Environmental IP66/NEMA 4 X case sealing
Approvals CE Mark
CSA; UL
Power supply 115/230 or 24 V ac ±15%. 50/60 Hz
Consumption < 40 VA (typical)
Product warranty 2 years parts and labor
PANEL METER, MODEL 21801
Display High-intensity, 7 segment, 1 x 6 digit
red LED display
Configuration User defined through keypad
Alarms Three user-defined
High/low process; High/low latch
Maths function Maximum and minimum value detection
Average value calculation
Analog inputs
Number One as standard
Input sampling rate 250ms per channel
Accuracy ±0.1 % of reading
Type Universal, configurable for:
Thermocouple, RTD, mV, mA, V dc
Linearizer functions Transmitter power supply
fitted as standard
Outputs
Retransmission Analog, configurable in range 4 to 20 mA
Relay output One 5A relay as standard (SPDT)
Assignable to alarms, totalizer count pulse, totalizer wrap
Environmental
Operating limits 32° to 131 °F (0° to 55° C)
5 to 95 %RH non-condensing
Temperature stability <0.02 % of reading at 1µV/°F (2µV/°C)
Front face NEMA 3/IP65
Approvals CE Mark
CSA; UL
Product warranty 2 years parts and labor
SENSORS AND FILTERS
HUMICAP
®
Sensors: Replacement HUMICAP
®
sensors are
available with and without membrane filters. The type of
replacement sensor recommended will depend on both the
instrument in use and its operating conditions. Refer to the
specific instrument data sheet for the appropriate replacement
sensor, or contact Vaisala for assistance.
Protective filters: Filter selection is important. Apart from
screening the sensor head from any stray electromagnetic
fields, the main purpose of the filter is to protect the sensor.
If the environment is clean and a particularly fast temperature
response is required, then an open cage grid can be used.
Otherwise a membrane or netting filter is recommended for
the majority of applications. When saturated, membrane or
netting filters dry out faster than sintered ones. For tempera-
tures above 176°F (80°C) or when there is a risk of damage
from fast moving particles, then a sintered or stainless steel net
filter should be used.
Vaisala probe heads can be installed in air streams with speeds
up to 40 meters per second when fitted with the appropriate
sintered filter.
Membrane filters:
Part # 2787HM ø 18.5mm, 0.5 µm
Part # 10159 ø 12.0mm, 0.5 µm
Filters for INTERCAP products:
Part# 15724 metallized plastic grid (non-filtering) ø 12 mm
Part# 16131 metallized membrane filter ø 12.0 mm, 0.5µm
Sintered bronze filters:
Sintered bronze filters are useful where mechanical strength is
necessary, in very dusty environments, or at extremely high
temperatures. Three options are available:
Part # 0195 HM ø 12 mm, 133 µm
Part # 6685 HM ø 18.5 mm, 37 µm
Part # 6686 HM ø 18.5 mm, 216 µm
Filters for the HMP 230 & HMP 360 series:
Part # 16452 sintered filter ø 13.5 mm, 38 µm
Part # 16720 PPS grid w/stainless steel net ø 13.5 mm, 14 µm
Part # 16562 PPS grid (non-filtering) ø 13.5 mm
Part # 16720 PPS grid w/membrane filter ø 13.5 mm, 2 µm
OTHER ACCESSORIES
11990 HM Mounting Flange: This stainless steel mounting
flange is easily attached to 12 mm diameter probes, allowing
them to be used with flanged connections.
90-2271 Power Supply: This power supply is UL-Approved,
uses 110 VAC power and supplies 24 VDC out.
85-HMPS Power Supply with NEMA 4 Enclosure
85-21797 Carrying Case: This case makes it easy to carry the
HMI 41, additional probes and other small items.
45-20618 Conduit Adaptor: Pg 9 to 1/2 in NPT
45-20775 Conduit Adaptor: Pg 11 to 1/2 in NPT
85-20557 Compression Fitting 14mm to 1/2 in NPT fitting
85-21804 Thread Adaptor 1/2 inch ISO to 1/2 in NPT
DMP248SC Sample Cell stainless steel (SS) chamber
DMP248BVS Ball Valve Assembly 1/2 in SS w/adaptor
TECHNICAL DATA - HUMICAP ACCESSORIES
BACK to
Table of
Contents
To INDEX
HUMICAP
Price List