Specifications

100 Commerce Way, Woburn, MA 01801
TEL: 1-888-VAISALA (824-7252) FAX: (781) 933-8029
E-MAIL: incsales@vaisala.com Access catalog on-line at: www.vaisala.com/inc/ssdcat
88
Note: Hectopascal (hPa) is the barometric pressure unit recom-
mended by WMO and also accepted by ISO.
1 hPa = 1 mbar
Operating Range
Pressure range
PTB 100A 800...1060 hPa
PTB 100B/PTB 101B 600...1060 hPa
PTB 101C 900...1100 hPa
Temperature range -40...+140°F (-40...+60°C)
Humidity range non-condensing
Accuracy
PTB100A/PTB101C PTB 100B/PTB101B
Linearity* ±0.25 hPa ±0.45 hPa
Hysteresis* ±0.03 hPa ±0.05 hPa
Repeatability* ±0.03 hPa ±0.05 hPa
Calibration uncertainty** ±0.15 hPa ±0.15 hPa
Accuracy at +68°F (+20°C***) ±0.3 hPa ±0.5 hPa
* Defined as ±2 standard deviation limits of end-point
non-linearity, hysteresis error or repeatability error
** Defined as ±2 standard deviation limits of inaccuracy
of the working standard at 1000 hPa including traceability
to NIST
*** Defined as the root sum of the squares (RSS) of end-
point non-linearity, hysteresis error, repeatability error
and calibration uncertainty at room temperature
Total accuracy
PTB 100A/PTB101C PTB100B/PTB101B
+68°F (+ 20 °C ) ±0.3 hPa ±0.5 hPa
0104°F (0...+40°C) ±1 hPa ±1.5 hPa
-4113°F (-20...+45°C) ±1.5 hPa ±2 hPa
-40140°F (-40...+60°C) ±2.5 hPa ±3 hPa
Long-term stability ±0.1 hPa/year
Effect of thermal or
mechanical shocks < ±0.2 hPa
General
Supply voltage 10 to 30 VDC
Supply voltage control with TTL level trigger when enabled
with an internal jumper, barometer can be triggered on/off
using external TTL level trigger
Supply voltage sensitivity less than 0.1 hPa
Current consumption less than 4 mA
less than 1 µA in shutdown mode
Output voltage
PTB 100A/PTB 100B 0...5 VDC
PTB 101B/PTB101C 0...2.5 VDC
Resolution 0.1 hPa
Load resistance 10 kohm minimum
Load capacitance 47 nF maximum
Settling time 1 s to reach full accuracy after
power-up
Response time (100% response) 300 ms
Warm-up shift less than 0.1 hPa
Acceleration sensitivity negligible
Pressure connector M5 (10-32) internal thread
Pressure fitting barbed fitting for 1/8" I.D. tubing
Maximum pressure limit 2000 hPa abs.
Electrical connector a removable connector for five
wires (AWG 28...16)
Housing material aluminum
Weight 85 g
BAROCAP
®
is a registered trademark of Vaisala. Specifications
subject to change without further notice.
PTB 100 Series
Dimensions in mm (inches)
8 7 .8 (3 .4 6)
7.2 (0.28)
11.5 (0.45)
7 6 .2 (3 .0 0)
4 (0.16)
3.5 (0.14)
60 (2.36)
21.8 (0.86)
13 (0.51)
9 6 .9 (3 .8 1 )
TECHNICAL DATA - PTB 100 SERIES
BACK to
Table of
Contents
To INDEX
PTB 100