User`s guide

User's Guide _______________________________________________________________________
44 ___________________________________________________________________ M210733EN-G
CAUTION
Take care not to damage the probe body. A damaged body makes the
probe less tight and may prevent it from going through the clasp nut.
CAUTION
In pressurized processes it is essential to tig
hten the supporting nuts and
screws very carefully to prevent loosening of the probe by the action of
pressure.
NOTE
Take care not to over tighten the clasp nut to avoid difficulties when
opening it.
Ball Valve Installation Kit for MMT338
The ball valve installation kit (Vaisala order code: BALLVALVE-1) is
preferred when connecting the probe to a pressurized process or pipeline.
Use the ball valve set or a 1/2" ball valve assembly with a ball hole of
ø14 mm or more. If you install the probe (ø12 mm) in a process pipe,
please note that the nominal size of the pipe must be at least 1 inch
(2.54 cm). Use the manual press handle to press the probe into the
pressurized (< 10 bar) process or pipeline.