User`s guide

USER'S GUIDE____________________________________________________________________
72 __________________________________________________________________ M211095EN-C
At the end of each command, press ENTER to execute the command. For
a successful execution, the following combinations of Carriage Return
<CR> and Line Feed <LF> are accepted:
- <CR>
- <LF>
- <CR><LF>
You can send commands at any Baud rate. The command is executed
when end-of-line character is received.
If messages have been enabled, WMT700 responds to an invalid
command with an error message. If the messages have been disabled,
WMT700 does not send any error messages. You can fetch the most
recent error data in configuration mode with the ERRORS command,
see section ERRORS — Get Error Codes and Counts on page 78.
NOTE
You can configure WMT700 by sending individual commands or by
loading a configuration file to the sensor. For instructions on configuring
all settings at once using configuration files, see section User-
Configurable Data Messages on page 83
Table 17 List of Configuration Mode Commands
Command Description
? Displays a list of configuration commands.
BAUD Changes or displays serial port settings.
CLEARERR Resets error counters.
CLOSE Switches the serial port to measurement mode.
ERRORS Fetches the error codes and counter information from WMT700.
G Displays either all or specified parameters.
H Displays list of data messages and available values for
measurement unit, profile, baud rate, interface, and analog
output mode.
MEAS Starts wind measurement based on the user-configurable
averaging time. WMT700 does not send data messages
automatically.
POLL Tests data polling.
RESET Resets WMT700.
S Changes selected parameters or defines new data messages.
START Starts continuous measurement.
STOP Stops continuous measurement.
VERSION Displays the software version.
WIND_GET Fetches wind calibration information.