User Manual

User's Guide ______________________________________________________________________
52 ______________________________________________________________________________
Calibration in progress CR LF
When the calibrate function completes successfully, the sensor response
will be the following:
Calibration successful, saving parameters CR LF
The sensor will save the parameters calculated in the non-volatile
memory. After the save completion, the sensor will perform a complete
power on the initialization sequence.
If the sensor returns the following response
Calibration failed CR LF
either the sensor or the zero wind speed verifier is broken. Issue the WS
command to determine if any self tests have detected a sensor failure
after removing the verifier and then waiting for 30 seconds. If the tests
fail, the sensor must be repaired.
WH Command
This command displays the sensor command Help menu. The command
format is:
WH CR LF
The WH command provides a list of all commands available for the
sensor, with a brief description of syntax, arguments (if any), and
description. This message is intended for display to a terminal
connected to the RS-232 sensor port.
Command/Response Summary CR LF
CR LF
WA Return average wind speed and direction CR LF
Response format:
CR LF
Byte 1: STX CR LF
Byte: 2 Sensor ID = W CR LF
Byte: 3 Command identifier CR LF
Byte: 4 Sensor Status: P = PASS, F = Failed,
H = Bad heater CR LF