Instructions Manual

Pub. #OM17-MSP3
4-30
Figure 75
2. Adjust electrode holder: with sampling mechanism raised completely, adjust electrode
holder to provide ½” (1.2 cm) clearance between it and sampling trough.
Figures 76a and 76b
3. Wash adjustment: Wash brackets should be parallel to sampling trough, with jets directly
beside electrode holder, jets should be 1” (2.5 cm) below electrodes; when electrodes are
properly aligned.
Figures 77a and 77b
Turn crank to raise
or lower electrode
holder
electrode
holder 1”
(2.5cm)
above wash
jets
Adjust scraper bracket
until cutting shoe
clears scraper when
sampler assembly is
raised completely.
½” (1.2 cm) clearance
between electrode holder
and trough liner