15. SENSE FUNCTION
The sense function is an automatic voltage control for the connection sockets MAIN OUTPUT (6).
In addition two separate measurement lines are connected parallel to the connection lines. The
voltage drop, which occurs at the connection lines, is measured on both these measurement
lines. This voltage drop automatically balances the laboratory power supply so that the voltage
actuallysettstheconsumer.
1. Switch off the power supply and consumer.
2. Alwaysrstconnectthepowersupplylinesofthepowersupplytotheconsumer.Please
note the correct polarity.
3. Press the terminal release inside with a small screw driver on the rear connection sockets
Remote Sensing (10) and stick the cables into the clamp openings. Check the secure
position.
4. Now connect both sense lines to the consumer at the correct polarity. The conductor cross
section for the sense lines must always amount to at least 0.34 mm
2.
5. Alwaysloosentheconnectionsinreverseorder(rstthesenselinesandthenthe
connection lines).
Remote
Sensing
Please ensure that the sense cables are connected as near as possible to
the connection point of the consumer. Please note the correct polarity.
Never short circuit the sense lines.