Instructions

Table Of Contents
Digital Storage Oscilloscope 15
[Run/Stop]: continuously acquire waveform or stop acquisition
[Single]: Acquire a single trigger, finish acquisition and then stop.
2.2 Connector
Figure 2-2 Connector
CH1, CH2, CH3, CH4: for an input connector of a measured signal.
EXT TRIG: be used as an input connector of an external trigger source. Use
[Trig Menu] button to select “External” trigger source (only for Edge trigger), and
the trigger signal source can be used for triggering in the third channel while
acquiring data.
Gen Out: Waveform signal output.
Probe compensation: The probe compensation signal is output and grounded
so that the probe is matched with the channels of the oscilloscope.
2.3 Universal Knobs and Softkeys
V0:
Universal Knob. Under different menu options, it supports selecting
menu options (
MEASURE), moving cursors and levels (Slope Trigger).
Press this knob to reset data (trigger holdoff
, overtime of the overtime
trigger and slope trigger), select menu options and so on. Easy to operate.
Press the
Wave Gen button on the front panel to open the arbitrary
waveform generator function.