Instructions

Table Of Contents
Digital Storage Oscilloscope 32
Lower is displayed at the upper right corner of the screen.
The Lower trigger level cannot be upper than the upper trigger level. V1 means the
upper trigger level while V2 means the lower trigger level.
2.7.7 Pattern Trigger
Identify a trigger condition by looking for a specified pattern. This pattern is a logical
"AND" or Or” combination of channels. Each channel can have a value of high (1), low
(0) or don't care (X). A rising, falling edge, rising or falling can be specified for one
channel included in the pattern. When an edge is specified, the oscilloscope will trigger
at the edge specified if the pattern set for the other channels are true (namely the
actual pattern of the channel is the same with the preset pattern). If no edge is
specified, the oscilloscope will trigger on the last edge that makes the pattern true. If all
the channels in the pattern are set to "Don’t Care", the oscilloscope will not trigger.
To set interval trigger:
1. Press the [Trig Menu] button on the front panel to enter the Trigger system function
menu.
2. Press the Type softkey, then use the Universal Knob to select Pattern and push
down the knob to confirm.
3. Press the Logic softkey, turn the Universal Knob to select a logical "AND" or Or
combination of channels and push down the knob to confirm.
4. Press Pattern to set the pattern of the current signal source, turn the Universal
Knob to select a pattern. At this point, the corresponding pattern is displayed on the