Digital Storage Oscilloscope 40
Sync Field The oscilloscope triggers when the synchronous field end.
Id Field The oscilloscope triggers when the Id field end.
Sync Id Error - The oscilloscope triggers when the Synchronization ID ERROR
end.
Identifier (Frame ID) - The oscilloscope triggers when a frame with an ID equal to
the selected value is detected. Use the Universal Knob to select the value for the
Frame ID.
ID and Data (Frame ID and Data) - The oscilloscope triggers when a frame with
an ID and data equal to the selected values is detected. Use the Universal Knob
to select the value for the ID and Data.
--Press Data software, use V0 to set the data, refer to
2.7.10;
--Data Mask: When set to "ON", the data is ignored when it is triggered; the
setting is "OFF", and the data on the data line must be consistent with the data
of the index so that it can trigger;
--Data Index: The range is 0 to 3. It can set four hexadecimal data.
For example, source: CH1; baud: 19200; Idle: High; When: ID and Data; ID: 0x25;
Data: 00 01 02 03. And set the trigger level. The trigger result is shown below: