Instructions

Table Of Contents
Digital Storage Oscilloscope 49
--Press Data software, use V0 to set the data, refer to 2.7.10;
--Data Mask: When set to "ON", the data is ignored when it is triggered; the
setting is "OFF", and the data on the data line must be consistent with the data
of the index so that it can trigger;
--Data Index: The range is 0 to 3. It can set four hexadecimal data.
Trigger Level:
5. When select SCL channel, press SCL and use Trigger Level knob to modify the
trigger level of the SCL channel. When select SDA channel, use Trigger Level knob to
modify the trigger level of the SDA channel.
For example1: SCL: CH1; SDA: CH2; When: Start Bit. The trigger result is shown
below:
For example2: SCL: CH1; SDA: CH2; When: Address and Data; Address: 09; Data: 3f
xx xx xx (The data mask of index 0 is OFF, the data mask of others are ON”). The
trigger result is shown below: