Digital Storage Oscilloscope 56
The measurement display area can display 4 measurement parameters at most, and
the measurements will arrange according to the selecting order. If add a sixth
measurement parameter, it will delete the first measurement.
Note: If the parameter does not match the measured condition, it will display as “*****”.
To Clear Measurement Parameters
Press the Clear All softkey to clear all the measurement parameters that are displaying
on the screen.
Statistic Function
Make statistic and display the current, average, minimum, maximum, root mean square
error and count values of at most 4 measurement items that are turned on last.
1. Press the [Meas] button on the front panel to enter the Measure function menu.
2. Press the Statistics softkey to select ON.
To Make All Measurement
All measurement could measure all parameters of the current measurement source
and display the results on the screen.
Do the following steps to make all parameters measurement.
1. Press the [Meas] button on the front panel to enter the MEASURE function menu.
2. Press the All Measure softkey to select ON.
3. Press the Source softkey to select the measure source (CH1~CH4).