YT format shows the vertical voltage in relation to
time (horizontal scale); XY format displays a dot
between CH1 and CH2 each time when a sample
is acquired, where the voltage or current of CH1
determines the X coordinate of the dot (horizontal)
and the voltage or current of CH2 determines the
Y coordinate (vertical). For detailed information,
refer to descriptions on XY format in the following
text.
Averages
4, 8, 16
, 32,
64, 128
Select the number of averages by pressing F3 or
F4.
Memory
Depth
Select
the memory depth for different board
models.
Normal: For the oscilloscope model with the bandwidth of 100MHz, the maximum
sample rate is 1GS/s. For time base with insufficient sample rate, you may use the Sine
Interpolation Algorithm to interpolate points between sampled points to produce a
complete waveform record (4K by default).
Peak Detect: Use this mode to detect glitches within 10ns and to limit the possibility of
aliasing. This mode is valid at the SEC/DIV setting of 4µs/div or slower. Once you set
the SEC/DIV setting to 4µs/div or faster, the acquisition mode will change to Normal
because the sample rate is fast enough that Peak Detect is unnecessary. The
Sample Points
Normal Acquisition Intervals
8
3
7
2
4
5
6
1
9
10
Normal Mode Acquires a Single Sample Point in Each Interval