Digital Storage Oscilloscope 64
Self Test Self test function.
Pass/Fail Pass/Fail function
Front Panel
Self Test
Test the function of all keys and knobs on the front panel.
Self Calibration: The self calibration routine can optimize the precision of the
oscilloscope to fit the ambient temperature. To maximize the precision, you should
perform the self calibration once the ambient temperature changes by 5 or more.
Follow the instructions on the screen.
Tip: Press any menu button on the front panel to remove the status display and
enter a corresponding menu.
2.12.1
Update Firmware
This series of oscilloscope can upgrade the software by the USB flash disc, which
needs about 5 minutes.
The firmware is upgraded by the following steps:
1. Plug the USB flash disc in which a firmware program is saved in USB Host interface
on the front panel of the oscilloscope.
2. Press [Utility] button to skip to the Utility menu.
3. Press Update->Update Firmware softkey.
4. Select the file and press Universal Knob to confirm. Then press Start updating
softkey to update the firmware.
Reboot the machine after finishing upgrading, and the software version is upgraded.
The oscilloscope should be self-corrected once after upgrading.
2.12.2
Self Calibration
The self calibration routine helps optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but should always run it if
the ambient temperature changes by 5 or more. For a more accurate calibration,
please power on the oscilloscope and wait for 20 minutes until it has adequately
warmed up.
To compensate the signal path, disconnect any probes or cables from the front-panel
input connectors. Then, push the [Utility] button, select Calibrate option and follow the