Digital Storage Oscilloscope 67
Functions Settings
Acquire Mode
Adjusted to Normal or Peak Detect
Cursor Off
Display Format Set to YT
Display Type Set to Vectors for an FFT spectrum; otherwise, unchanged
Horizontal Position
Adjusted
SEC/DIV
Adjusted
Trigger Coupling
Adjusted to DC, Noise Reject or HF Reject
Trigger Holdoff
Minimum
Trigger Level
Set to 50%
Trigger Mode
Auto
Trigger Source
Adjusted; Auto Scale can not be used for the EXT TRIG
signal
Trigger Slope
Adjusted
Trigger Type
Edge
Trigger Video Sync
Adjusted
Trigger Video
Standard
Adjusted
Vertical Bandwidth
Full
Vertical Coupling
DC (if GND was chosen before); AC for the video signal;
otherwise, unchanged
VOLTS/DIV
Adjusted
The Auto Scale function examines all channels for signals and displays corresponding
waveforms. Auto Scale determines the trigger source according to the following
conditions.
If multiply channels get signals, the oscilloscope will use the channel with
the lowest frequency signal as the trigger source.
If no signals are found, the oscilloscope will use the lowest-numbered
channel displayed in Auto Scale as the trigger source.
If no signals are found and no channels are displayed, the oscilloscope will
display and use Channel 1 as the trigger source.
When you use the Auto Scale function and the oscilloscope determines that the signal
is similar to a sine wave, the oscilloscope displays the following options.
Wave Options Details
Multi-Period
Display multiple periods that have appropriate vertical and
horizontal scales.