User Guide

use your target bare metal substrate instead of our
attached substrates.
Note: Once you did a misoperation, please
press and hold MODE/RESETKEY 3 seconds
for resetting the factory default.
ZERO Calibration: Press "MODE" to the Single
Measurement Mode(SNG in LCD), the instrument
measures the attached iron-based, aluminum-based
or uncoated bare metal substrate, measures several
times to get stable reading; The next step is to ZERO
Calibration, long press CALIB/ZERO key until the
buzzer beeping three times (bebe..be), now ZERO
icon is display in the LCD and Reading is 000.0, then
measures substrates to do a check. If any error occurs
in ZERO Calibration, Please long press MODE/RESET
to do system resetting and ZERO Calibration again.
(Hint to Plating measurement customers: For the
plating layer is generally thin, the best way is to do
ZERO Calibration on the your target unplated bare
metal layer before measurement to ensure accurate
reading.)