2.7 Probe Mode
There are three probe modes: AUTO, MAG and EDDY. In AUTO mode, the probe can automatically determine
the type of substrate measured. In MAG mode, the probe can only measure on magnetic substrates. In EDDY
mode, the probe can only measure on non-ferrous metal substrates. When a magnetic substrate detected, an
“Fwill be shown on the right of readings, and when a non-ferrous metal detected, a “Nwill be shown. User
can switch probe mode as below.
a. Press Left Keyonce to enter into menu mode(Rootdirectory);
b. Press “Up Keyor Down Keyto activate Optionsitem, and press Left Keyonce to enter in;
c. Press Up Keyor Down Keyto activate ”Probe Modeitem, and press Left Keyonce to enter in;
d. Press “Up Keyor Down Keyto activate your intended item, and press Left Keyonce to select the item
and go back. Press ZERO Keyto back measure mode.
2
.
8 Storage
In general group mode(GENn), readings and statistics will be stored and not be lost when power off. In direct
group mode(DIR), readings and statistics will be lost when power off.
Each group has individual high/low limit alarm, zero calibration and multi-point calibration, user can change
these settings, and they will be stored.
In addition, user can set the system settings(e.g.: Measure Mode,Group Mode,Probe Mode etc.), these settings
will also be stored.
Note
:
When
battery
low, user must
replace
the
battery timely. B efore replacing
the
battery,
you need to power off
the gauge firstly.
2.9 Recall and Delete Readings
Delete the last readings(Tips: If in measure mode, press the “UP Keyonce by one beep).
a. Press Left Keyonce to enter into menu mode(Rootdirectory);
12