User Manual

4.1 Zero Calibration
Tips: For P3 or P4, press and hold DOW N Keyto start zero calibration, and press again to exit
a. Press and hold ZERO Keyuntil hearing beep...beep beep...”. The LCD will flash ZEROsymbol.
b. Place the probe on uncoated sample and raise it after one beep.
c. Repeat step b and c several times to obtain mean value.
d. Finished. Press ZERO Keyto exit calibration mode.
4
.
2 Multi-point Calibration
Tips: For P3 or P4, user needs to access the
menu to enable or disable multi-point cal:
Press Left key->Select Calibration->Select
Point Cal->Select Enableor Disable
a. In measure mode, press CAL keyonce to enter into calibration mode.
b. Zero calibration according to chapter 4.1 and then lay the calibration foil on an uncoated sample.
c. Apply the probe and raise it after one beep. a readings will be shown.Press “Up Keyor Down Keyto
increase or decrease the readings or hold the key for changing continuously until the required foil thickness
value shown.The LCD will show current calibration point Ptn X . The “Xwill be flashing.
d. Apply the probe to the test sample several times, and every time raise the probe, if press CAL Key, after
beep...beep beep...”, it will establish a mean value for current calibration point.
e. Press Left Keyto confirm and end current calibration point, and the “Xwill be steady. Or press Right
Keyto cancel and end current calibration point.
f. For more calibration points, use another calibration foil and repeat the step c, d and e.
g. Finished. Press CAL Keyto exit calibration mode.
Note:
a. Each group has individual zero calibration and multi-point calibration .
b. When calibration up to 4 points, user must clear the finished points firstly to restart.
c. Readings measured before will not be affected by new calibration.
d. In calibration mode, measuring will not be affected.
e. Recommended to work in single measure mode when calibrating.