Instruction Manual

17
4.11 SCALING AND CALIBRATION
Since a computer is used to display the reading and load the setpoints, a mathematical
step can be used to convert measurements and setpoints to engineering units and correct
for known sensor calibration errors.
For example, the ANAFAZE 8 PID does all thermocouple calculations in degrees F since
this provides almost twice the resolution of degrees C. If degrees C display and setpoints
are desired the computer makes the F to C conversion as data is received from the
ANAFAZE 8 PID and converts the setpoints from C to F as they are sent to the
controller.
In a similar manner, linear sensors can be converted to engineering units and adjusted for
known calibration errors with a conversion step. For a linear sensor two outputs can be
measured (x1 and x2) and converted into engineering units (y1 and y2) using the
standard formula:
y = mx + b
where m = (y2 - y1)/(x2 - x1)
and b = y2 - mx2 or b = y1 - mx1\
The same conversion formula can be used to convert the desired setpoint into a
percentage of full scale which allows the ANAFAZE 8 PID to control to the actual
measured signals while the computer displays the measurements and setpoints in
engineering units. This approach eliminates the need for potentiometers and other analog
adjustments on each input channel.
The ANASOFT PID software for the IBM PC and compatible computers includes these
scaling functions as part of the menu driven program. Please consult ANAFAZE for
additional information.
4.12 RTD INPUTS
RTD's can be connected in different configurations including bridge circuits, three wire
and four wire -- please request a copy of the ANAFAZE RTD application bulletin.