Instruction Manual

3
The user may select an alarm deadband of 0 to 9 % of the setpoint. Refer to the
command summary at the end of this section for further information.
Entering a deviation percentage of 0% (which is the default value) will result in no
deadband and alarms will be set and cleared as measured temperatures cross the
alarm setpoints.
NOTE:
Entering a new deviation percentage will result in the 8 upper and 8
lower deadband limits being recalculated instantly. Entering a new alarm
setpoint (high or low) for any loop will result in the deadband limit for that
alarm setting being recalculated at that time.
1.1.3 GLOBAL ALARMS
Two global alarms are provided in Automatic Alarm Mode: one for high alarms and
one for low alarms.
The alarm output AL1 on the PID is set if any one of the 8 Expander Card High
Alarm Outputs is on. It is cleared when no High Alarm Outputs are Set on the
Expander Card.
The alarm output AL2 on the PID is set if any one of the 8 Expander Card Low
Alarm Outputs is on. It is cleared when no Low Alarm Outputs are Set on the
Expander Card.
1.2 I/O CONTROL MODE
In I/O control mode lines 0 to 15 are under user control and may be defined as inputs or
outputs and written or read accordingly.
Refer to the command summary at the end of this section for further information
regarding commands to set, clear and interrogate the status of I/O lines.