Propaq LT Vital Signs Monitor ® HALL, ROBERT E. 3456187 Adult II SpO2 3:00:06P Rm 239 1mV/cm 2x 80 140/78 12 97 % NIBP mmHg (102) Resp/min SpO2 HR/min @2:47P Manual Directions for Use 901061 Patient monitor Software version 1.60.
ii Welch Allyn Propaq LT Vital Signs Monitor © 2014 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn.
iii Contents 1 - Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
iv Contents Welch Allyn Propaq LT Vital Signs Monitor About ParamSet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 Responding to an alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 Responding to an alert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 About battery charge status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 1 Introduction Intended use The Propaq LT Series (802LTAN, 802LTAS, 802LT0N, 802LTRN, 802LT0S, and 802LTRS) monitors are portable devices intended to be used by clinicians and medically qualified personnel for single- or multiparameter vital-signs monitoring of ambulatory and nonambulatory patients, including neonate, pediatric and adult patients. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2.
2 Introduction Welch Allyn Propaq LT Vital Signs Monitor Table 2. Control buttons Monitor Power - Turn the monitor power on or off. Down - Move the cursor down to the next display item. Action - Act based on what is highlighted. (See “Using the action button” on page 31.) Silence/Reset - Silence the current alarm tone for 90 seconds or reset a silenced alarm tone. Left - Move the cursor left to the next display item; decrease the parameter value.
Directions for Use Introduction Table 3. Status indicators (continued) (red) The battery is near failure; the monitor will shut down soon. If this indicator appears while the monitor is in the cradle, the battery cannot be charged and must be replaced. The battery is near failure and is charging. The monitor will shut down if removed from the cradle. The monitor is communicating wirelessly with the network and with Acuity. (Wireless only, Acuity enabled.
4 Introduction Welch Allyn Propaq LT Vital Signs Monitor Table 4. Labels High voltage. Do not touch during defibrillation. Direct current. Recycle the battery separately from other disposables. Li ++ Battery replacement specification. Lithium-ion battery. This monitor (2.4 GHz) is approved for use during all phases of flight aboard U.S. Army aircraft. AWR: MIL STD: 461E. Army: CE101, CS101, T, CS115, S116, RE102, RE103. Rechargeable battery Cradle Power in (DC). Fuse replacement specification.
Directions for Use Introduction 5 General warnings These statements apply to all aspects of patient monitoring. Statements which apply specifically to one aspect of monitoring, such as NIBP or SpO2 monitoring, are presented in the corresponding sections of the manual. WARNING Many environmental variables, including patient physiology and clinical application, can affect the accuracy and performance of the monitor. The clinician must verify all vital-signs information prior to patient intervention.
6 Introduction Welch Allyn Propaq LT Vital Signs Monitor WARNING To comply with Federal Communications Commission (FCC) RF exposure requirements and to avoid exposure to radio-frequency (RF) radiation, always use the monitor in accordance with the operating conditions and instructions provided in this manual. WARNING Pacemaker signals can differ from one pacemaker to the next.
Directions for Use Introduction 7 WARNING This wireless medical device was tested and, when used with a metal-free accessory between the monitor and the patient, complies with FCC RF Exposure (SAR) guidelines. The use of accessories containing metal may not ensure compliance with FCC RF exposure guidelines. Specific Absorption Rate (SAR) is a measurement of radio frequency energy. The FCC permits a maximum SAR value of 1.6 mW/g.
8 Introduction Welch Allyn Propaq LT Vital Signs Monitor Controls, indicators, and connectors Figure 1. Controls HALL, ROBERT E. 3456187 Adult, Display Cycle to the next configured display format, or cancel the current control, setup, or pop-up menu. Power Turn the monitor power on or off. Silence/Reset Silence the current alarm tone for 90 seconds or reset a silenced alarm tone. Move the display cursor up, down, right, or left to highlight an item; change parameter values.
Directions for Use Introduction HALL, ROBERT E. 3456187 Adult, 3:00:06P Rm 239 II 1mV/cm Figure 3. Indicators: cradle SpO2 2x 40/78 78 12 97 80 140 97% NIBP mmHg (102) Resp/min SpO2 HR/min @2:47P Manual HALL, ROBERT E.
10 Introduction Welch Allyn Propaq LT Vital Signs Monitor Features and functions • • • Monitoring of neonate, pediatric and adult patients Display of ECG, SpO2, and Resp waveform traces Accurate reading of NIBP in the presence of motion artifact, using Welch Allyn’s ® patented Smartcuf motion-tolerant technology SpO2 monitoring with advanced technology for accuracy under conditions of low perfusion ™ Configurable adjustments to alarm limits with ParamSet technology Standalone operation with local patien
Directions for Use Introduction 11 Accessories The following accessories are available for use with the monitor and the cradle: • • • • • • • • • • • • • • Large Color Display Interface and cables Propaq LT Monitor PC Utility software (CD) Propaq LT Monitor Service Manual (CD) Propaq LT Monitor Directions for Use (CD) Patient carry strap Patient wearable strap Transport stretcher strap Connector panel plugs SpO2 cables and sensors 3-lead and 5-lead ECG cables and cable extensions ECG electrodes NIBP hos
12 Introduction Welch Allyn Propaq LT Vital Signs Monitor HIPAA considerations Each medical facility is responsible for creating and enforcing policies and procedures to guarantee compliance with the regulations defined in 45 CFR 160-164 of the Health Insurance Portability and Accountability Act (HIPAA) of 1996.
13 2 Overview of monitor operation Turning on the monitor The monitor runs through an operational self-test each time it is powered on. Always verify that it follows the power-on sequence shown below. When Step 3 completes as described, the monitor has verified that the visual and audible alarm indicators are working properly. If Step 4 does not complete as described, remove the monitor from service and have it examined by a qualified service person. To turn on the monitor 1.
14 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor Selecting a language To change the language of the monitor interface 1. If the monitor power is on, press 2. Press to turn it off. to turn the monitor power on. 3. Immediately after pressing , and before any screen appears, simultaneously press and . Keep the buttons pressed until the language selection screen appears. Figure 8. Language selection screen English Deutsch Français Español Italiano 4. Press , 5. Press .
Directions for Use Overview of monitor operation 15 WARNING Never download a configuration file to the monitor while you are monitoring a patient with the monitor in the cradle. • Communication with the network (and Acuity) is interrupted if you download a configuration to the monitor or upload patient data from the monitor. • Changing the monitor configuration shuts down the monitor and deletes all patient data from the monitor. Figure 9. Monitor and cradle Monitor HALL, ROBERT E.
16 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor 2. Connect the AC adapter to an AC power outlet and to the cradle. The indicator (green) on the front of the cradle indicates that the AC power adapter is connected. Figure 10. Cradle status indicators HALL, ROBERT E.
Directions for Use Overview of monitor operation Displaying data Patient vital signs appear on the 3.5-inch (diagonal measurement) monitor display and optionally on the Large Color Display screen. Figure 12. Components of the vital-signs display (A) 1 2 3 4 HALL, ROBERT E.
18 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor Figure 13. Components of the vital-signs display (B) HALL, ROBERT E. 3456187 Adult II 3:00:06P Rm 239 1mV/cm 9 SpO2 10 2x 12 97 % 80 140/78 NIBP mmHg (102) Resp/min SpO2 HR/min @2:47P Manual 15 9 10 14 11 13 12 Battery status (See Table 3, “Status indicators” on page 2.
Directions for Use Overview of monitor operation 19 About display formats The monitor can be configured to display any of these formats: Table 5. Display formats HALL, ROBERT E. 3456187 Adult 3:00:06P Rm 239 % 80 97 140/78 78 12 HR/min SpO2 NIBP mmHg (102) Manual 2:47P S D Blood pressure, heart rate, respiration rate, and SpO2 measurements displayed in large numerics. Bell icons indicating... Left half Right half White Black Resp/min M HALL, ROBERT E.
20 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor The monitor can be configured to cycle quickly through three of the five available display formats when you press . For example, if the configuration specifies three formats— Large Numerics, Single Waveform, and Tabular Trends—you can press repeatedly to cycle through those formats. Figure 14. Cycling through the configured display formats HALL, ROBERT E.
Directions for Use Overview of monitor operation 21 Locking out the display, back light, and buttons If lock-outs are enabled, you can lock out one or more of the following elements: Buttons To prevent unauthorized use Back Light To extend the battery run life Display To prevent unauthorized viewing of patient information The monitor configuration defines which of these, if any, can be locked out.
22 Overview of monitor operation Note Welch Allyn Propaq LT Vital Signs Monitor If the interface box is attached to the cradle, the bed rail hook cannot be used. Figure 15.
Directions for Use Overview of monitor operation Figure 16. Interface box mounted on the cradle Interface box mounted on the cradle Power conditioner Figure 17.
24 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor Large display viewing options The large display shows vital signs numerics and either four or nine waveforms. Figure 18.
Directions for Use Overview of monitor operation 25 Leads I, II, and III are viewable. The waveform for the lead selected at the monitor is displayed on the large color display. Lead V cannot be displayed. Note When using a 5-lead cable with all electrodes attached, only leads II and V are available in the 4-waveform view. Figure 21. 9-waveform view, 3-lead cable Leads I, II, and III are viewable. The waveform for the lead selected at the monitor is displayed on the large color display.
26 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor When the interface box is not used, the AC power adapter is attached to the cradle. When the interface box is used, the AC power adapter is connected to it and not to the cradle.
Directions for Use Overview of monitor operation 27 Detaching the bed rail hook from the cradle If you are attaching the interface box to the cradle and if the cradle and the bed rail hook are attached, you must first detach the bed rail hook from the cradle. 1. Remove the screws from the back of the bed rail hook. 2. Lift the bed rail hook off of the shoulder screws on the back of the cradle. 3. Remove the shoulder screws from the back of the cradle. Figure 23.
28 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor Assembling the large color display Refer to Figure 24, Figure 25 on page 29, or Figure 26 on page 30. Figure 24.
Directions for Use Overview of monitor operation Figure 25.
30 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor Figure 26.
Directions for Use Overview of monitor operation 31 About navigation You navigate the monitor screens using and (display button). , , and (arrow buttons), (action button), Using the arrow buttons Use , , and to do the following: • Highlight an item on the display. (See “Using the highlights” on page 31.) • Select options from a control menu. • Use • Use and and to select options from a pop-up menu. to change the values of numeric parameters.
32 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor Figure 27. Examples of highlighted elements HALL, ROBERT E. 3456187 Adult II 12:41:32 Rm 239 Drop-down menu 1mV/cm Tabular Time 12:41 12:40 12:39 12:38 12:38 12:37 12:36 12 SEARCH % Resp/min SpO2 80 140/78 HR/min NIBP mmHg 125 122 100 75 50 25 130/65 (93) n/a 112/87 (87) n/a n/a 192/110 (130) n/a n/a 22 18 16 17 19 19 98 99 98 99 100 n/a 98 Highlights Waveform Size 0.2 mV/cm 0.
Directions for Use Overview of monitor operation 33 Menus Using control menus Figure 29. SpO2 control menu (example) HALL, ROBERT E.
34 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor 2. Scroll (using ) to HR/PR Tone, and press either highlight Off. Note as many times as needed to When you change a setting (for example, by turning off an alarm limit or by increasing or decreasing an alarm limit), the change takes effect immediately. 3. Press Note or or to exit the control screen and return to the vital-signs display.
Directions for Use Overview of monitor operation 35 About the NIBP control menu The NIBP control has submenus Manometer, Systolic, Diastolic, and Mean. Press to select one. or Figure 30.
36 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor About the Resp control menu Parameter Options Parameter Options Resp Monitoring Off On Lower Alarm Off On Upper Alarm Off On Lower Limit Upper Limit Adult 2 - 148 Adult 4 - 150 Pediatric 2 - 148 Pediatric 4 - 150 Neonate 3 - 148 Neonate 5 - 150 Resp Lead Ld1 (RA-LA) Ld2 (RA-LL) Using setup menus Use the Setup menus—Alarms, ECG, NIBP, SpO2, and Timings—to define settings for monitor behavior.
Directions for Use Overview of monitor operation 37 Use the Setup -> Configuration menu to change the monitor configuration without using the Configuration Utility. Option Note Configuration Select ( ) to save current settings. Replace the power-up defaults with the current parameter settings. Select ( ) to use factory settings. Replace the power-up defaults with the factory settings. Select ( ) to use file: [file name] Replace the power-up defaults with the last downloaded configuration settings.
38 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor About monitor information screens Both the start-up information screen and the monitoring information screen provide information about the monitor. To view the start-up information screen From the power-on screen (Figure 7 on page 13), highlight Info. Figure 32. Start-Up Information Welch Allyn Propaq 802LTRN Serial# F82C0DD5 V1.00.00 Portland Westside Emergency Department PtldWstsdED10Jun05.
Directions for Use Overview of monitor operation 39 To view the monitoring information screen Highlight (in the upper right corner of the vital-signs display) and press . Figure 33. Monitoring information HALL, ROBERT E. 3456187 Adult II 12:41:32 Rm 239 1mV/cm 80 (100) 12 HR/min NIBP mmHg (148-86) Resp/min Portland Westside Emergency Department PtldWstsdED10Jun05.
40 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor Using demo mode In Demo mode, the monitor displays simulated patient data for all vital signs. You can use Demo mode to familiarize yourself with the following: • Viewing vital-signs displays • Modifying alarm limits and other settings • Cycling through display formats • Responding to alarm conditions To start Demo mode, all of the following must be true: • The power-on screen is displayed.
Directions for Use Overview of monitor operation 41 Figure 35. Demo mode: initial display ID: 01018VDO9PBH SIMULATION 3:00:06P Adult Rm 239 % 80 97 12 HR/min SpO2 NIBP mmHg S D Resp/min M When you enter Demo mode, the monitor is in ‘Demo mode low’. In this mode, the simulated vital signs of the patient are steady and do not cause any alarms at the default alarm limit settings.
42 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor The Timings menu specifies the Demo mode—Demo mode low, which simulates normal vital signs, and Demo mode high, which simulates a higher heart rate (HR), higher respiration rate (Resp), and lower oxygen saturation (SpO2). 3. Press to highlight Demo mode, and press or to highlight High. Figure 37.
Directions for Use Overview of monitor operation Figure 39. Demo mode high: simulated alarm condition Flashing red: patient alarm HR limit violation (red) ID: 01018VDO9PBH HALL, E.
44 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor Power saving To maximize battery life, the monitor display shuts off when the following conditions are all true: • No button press for a period of n seconds. (n is configurable.) See “Timing out the display and the back light” on page 20. • No active alarms or alerts • No Acuity Message window • No Patient ID entry window Turning off the monitor To turn off the monitor Press . The Power-off screen appears: Figure 41.
Directions for Use Overview of monitor operation 45 Transporting the monitor with the patient An ambulatory patient can wear or carry the monitor using the wearable strap or the patient carry strap (optional accessories). Medical personnel can use the transport stretcher carry strap to keep the monitor with a patient during stretcher transport.
46 Overview of monitor operation Welch Allyn Propaq LT Vital Signs Monitor To use the patient carry strap WARNING Do not use the patient carry strap to lift or carry both the monitor and the cradle. The patient carry strap is not intended to support (and cannot support) the weight of both the monitor and the cradle. Attempting to carry both the monitor and the cradle with a patient carry strap could lead to patient injury and to damage to the monitor and the cradle. 1. Remove the monitor from the cradle.
47 3 Standalone monitoring Overview The monitor is available in standalone models (802LT0N and 802LT0S) and in wireless models (802LTAN, 802LTAS, 802LTRN and 802LTRS). This chapter describes the operation of the standalone monitor and of the wireless monitor when it is not in communication with the Welch Allyn Flexnet® network and an Acuity® Central Station. About the Model 802LT0N and 802LT0S (standalone) monitors The model 802LT0N and 802LT0S monitors operate in standalone mode.
48 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor Figure 45. Start-up with saved data Portland Westside Hospital Emergency Department Note! There is patient data stored for Hall, Robert E. ID: 3456187 Select ( ) to delete data and start a new patient. Start New Patient Continue Patient • Info Demo If the data from the previously monitored patient was deleted on shut-down, the “no data saved” start-up screen appears: Figure 46.
Directions for Use Standalone monitoring 49 Figure 48. Patient information entry Last Middle First ID# Rm# Entry fields A B C D E F G H I J K L M N O P Q R S T U V W X Y Z 1 2 3 4 5 6 7 8 9 0 - Backspace Character field Select ( ) to enter highlighted character Confirm Cancel 5. Enter the ID and name of the new patient. a. Press b. Press c. to highlight (blue) a character in the character field (Figure 49). to copy it to the green-highlighted location in the entry fields (Figure 48).
50 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor Figure 50. Changing patient mode Patient Mode Adult Pediatric Neonate Patient Mode Adult Pediatric Neonate c. A confirmation screen appears: Figure 51. Confirming a change of patient mode HALL, ROBERT E. 3456187 Adult 3:00:06P Rm 239 Note! You are changing the patient mode from Adult to Pediatric. % This will change settings to default Pediatric values and delete previous HR/min SpO2 patient data.
Directions for Use Standalone monitoring 51 Figure 53. Time/date screen: 12-hour format HALL, ROBERT E. 3456187 Adult II 14:49:54 Rm 239 1mV/cm HR/min NIBP mmHg Resp/min % SpO2 Time Format 12 Hour 24 Hour 2P Hour Minute 45 Month 3 Day 17 Year 2005 Date Format mm/dd/yy dd.mm.yy yy/mm/dd Set Cancel c. Note 12-hour format When the time and date are correct and formatted appropriately, press accept the changes and return to the vital-signs display.
52 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor Figure 55. Start-up with saved data Portland Westside Hospital Emergency Department Note! There is patient data stored for Hall, Robert E. ID: 3456187 Select ( ) to continue monitoring this patient. Start New Patient Continue Patient Info Demo Monitoring ECG and Resp Overview Note In this manual, lead refers to an ECG electrode or lead wire, and Lead refers to a waveform source.
Directions for Use Standalone monitoring 53 WARNING (1) During a surgical procedure, do not use small ECG electrodes. (2) Select ECG electrode attachment points remote from the surgical site and remote from the electrosurgical return electrode. (3) Use electrosurgical return electrodes with the largest practical contact area. (4) Assure proper application of the electrosurgical return electrode to the patient.
54 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor Monitoring ECG 1. Inspect the ECG cable. Replace it if it shows any signs of wear, breakage, or fraying. 2. Plug the cable into the monitor. 3. Select electrode sites on the patient, choosing flat areas and avoiding fatty or bony areas and major muscles. Figure 56. ECG leads - actual placement RA LA V1 V6 Six possible V lead electrode placement sites for the C lead.
Directions for Use Standalone monitoring 55 Figure 57. ECG leads - placement displays, 5-lead and 3-lead The locations of the circles displayed on the monitor (Figure 57) for each lead are fixed, and do not indicate the exact placement of the electrodes on the patient. If the monitor detects that some lead wires are not connected, it displays an “ECG Fault” equipment alert and a chest diagram indicating the location of the disconnected lead or leads.
56 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor Figure 60. Waveform source: II STEWART, ANN 7762940 Adult II 1mV/cm Waveform Source Lead I Lead II 6 sec Lead III aVR aVL aVF V Resp SpO2 Add 2nd Wave HR/min NIBP mmHg 80 15:14:28 Rm 263 12 97 % Resp/min SpO2 2. Highlight your waveform source choice and press or . Figure 61.
Directions for Use Standalone monitoring 57 About pacemakers and ECG monitoring If the patient being monitored has a pacemaker, the monitor detects and can indicate the occurrence of pacemaker signals. If the Pacer Indicator setting is ON, the monitor displays and prints vertical dashed lines to indicate detected pacemaker signals. If Pacer Indicator is OFF, the monitor continues to detect the pacemaker signals but does not display or print the pacer markers. WARNING Signals differ between pacemakers.
58 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor Improving the waveform display If the power source filter is off, noise from the power source can cause an unclear or noisy waveform. To change the power source filter to reduce noise 1. Access the Setup menu. (See “To access the setup menus” on page 37.) 2. Highlight ECG. Verify that the setting for Power source filter is correct for the power source in your facility.
Directions for Use Standalone monitoring 59 Figure 65. Resp waveform, default size (2x) HALL, ROBERT E. 3456187 Adult Resp 3:08:16 Rm 239 2x 80 HR/min 12 97 % NIBP mmHg Resp/min SpO2 For more detail, change the waveform size to 8x. (See “To change the waveform size” on page 56.) Note Waveform size does not affect breath-detector sensitivity. Figure 66. Resp waveform enlarged for detail (8x) HALL, ROBERT E.
60 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor Monitoring SpO2 The monitor can be purchased with the Nellcor or the Masimo SpO2 option. Most of this chapter pertains to either configuration. Monitors with the Masimo technology, however, can provide additional SpO2 data. (“About the Masimo SpO2 display” on page 64.) WARNING Accurate measurements of oxygen saturation, when derived from pulse oximetry, depend to a great extent on patient condition and proper sensor placement.
Directions for Use Standalone monitoring 61 WARNING Interfering substances: Carboxyhemoglobin can erroneously increase readings; the level of increase is approximately equal to the amount of carboxyhemoglobin present. Methemoglobin and other dysfunctional hemoglobins can also cause erroneous readings. Further assessment beyond pulse oximetry is recommended. Intravascular dyes, or any substances containing dyes, that change usual arterial pigmentation can cause erroneous readings.
62 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor - the patient is in cardiac arrest or shock the patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia • If poor perfusion affects performance for an adult, consider using the adult nasal sensor.
Directions for Use Standalone monitoring 63 Figure 68. SpO2 monitoring turned off HALL, ROBERT E. 3456187 Adult 3:11:24P Rm 239 % 80 HR/min SpO2 Monitoring is Off SpO2 @ 3:03P NIBP mmHg Resp/min S D M To take an SpO2 spot check reading 1. Enable spot checks. (See “To prepare to take a spot check reading” on page 62.) 2. Attach the sensor to the monitor and the patient. 3. Highlight SpO2 @ XX:XX and press Note . The SpO2 drop-down menu appears.
64 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor Figure 71. SpO2 spot check: pulse rate reading HALL, ROBERT E. 3456187 Adult 3:12:13P Rm 239 78 98 PR/min % SpO2 @ 3:05P Pulse rate NIBP mmHg Resp/min S D M • The spot check ends, and SpO2 monitoring is again turned off. • The SpO2 text on the display screen now includes the time of the most recent SpO2 measurement. For example: SpO2 @ 3:05P. Note Spot-check readings are included in trend displays. 5.
Directions for Use Standalone monitoring Normal For typical monitoring. APOD For monitoring when the probe is likely to detach from the patient because of wet skin, motion, or other unusual circumstances. Maximum For monitoring patients with extremely low perfusion. Averaging Time 65 The period (in seconds) over which multiple SpO2 measurements are taken and a result is derived.
66 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor To monitor blood pressure 1. Select cuff size based on limb circumference. Use only hoses and cuffs listed in the accessory list or at www.welchallyn.com. 2. Squeeze all the air from the cuff before placing the cuff on the patient. 3. Place the cuff on the limb, as near heart level as possible. Figure 72. Cuff placement Cuff applied evenly and snugly. The center of the cuff is at heart level, and the bottom edge is one inch (2.
Directions for Use Standalone monitoring 67 Figure 74. NIBP: reading in progress STEWART, ANN 7762940 NIBP in progress. II 1mV/cm HR/min Adult NIBP mmHg NIBP Manometer Resp/min Systolic 3:00:06P Rm 263 Diastolic SpO2 % Mean Manometer pressure indicator bar NIBP Mode: Manual 145 50 Exit 100 Trends 150 200 Snapshots 250 300 Setup When the cycle completes, control returns to the previous direct-access display.
68 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor Note The message ???/???/(???) in an NIBP TREND display or printout indicates that the monitor could not complete an NIBP measurement during that period. Note On-demand NIBP readings (manual or turbo) are delayed by 8 seconds if they are started while the monitor is in a power-saving mode such as display time-out.
Directions for Use Standalone monitoring 69 To stop automatic NIBP readings 1. Highlight NIBP and press . 2. Highlight NIBP Mode and select Manual. Taking NIBP readings using Turbo mode In Turbo mode, the monitor starts an NIBP reading and then takes as many more readings as possible within five minutes. To use the Turbo mode 1. Select and apply the NIBP cuff and hose. (“Monitoring blood pressure (NIBP)” on page 65.) 2. Highlight NIBP and press . 3. Highlight NIBP Mode and select Turbo.
70 Standalone monitoring • Welch Allyn Propaq LT Vital Signs Monitor Download another configuration from a PC. (See “Using the Configuration Utility to configure the monitor” on page 116.) ID: 01018VDO9PBH 3:00:06P To save the current settings Adult as Rmthe 239 new default settings II 1mV/cm 1. Access the Configuration Setup menu. 80 HR/min 12 Resp/min NIBP mmHg 97 % SpO2 Setup Alarms ECG NIBP Timings Configuration Service Select ( ) to save current settings.
Directions for Use Standalone monitoring ID: 12549M6I9ERF 71 03:53:45 Adult Rm Note! You are changing the default settings to factory settings. % If this is correct, highlight HR/min SpO2 “Confirm” and then press Select ( ). Cancel Confirm NIBP mmHg S D M To discontinue monitoring 1. Press . The Power Off screen appears. Figure 76. Power-off HALL, ROBERT E. 3456187 Adult 15:01:24 Rm 239 You have pressed the Power Off button. There is patient data stored for Hall, Robert E.
72 Standalone monitoring Welch Allyn Propaq LT Vital Signs Monitor
73 4 Monitoring in communication with Acuity About wireless monitoring The model 802LTAN, 802LTAS, 802LTRN and 802LTRS wireless monitors can operate either in standalone mode or in communication with a Welch Allyn Acuity Central Station. (See “Standalone monitoring” on page 47.) While the wireless monitor is connected to the FlexNet network, patient data gathered by the monitor is continuously stored at Acuity.
74 Monitoring in communication with Acuity Note Welch Allyn Propaq LT Vital Signs Monitor When the monitor is used in Canada: To provide maximum shielding (to prevent radio interference to the licensed service), the monitor is intended to be operated indoors and away from windows. Radio equipment (or its transmit antenna) that is installed outdoors is subject to licensing.
Directions for Use Monitoring in communication with Acuity 75 Monitoring a patient outside of network range Monitors are typically deployed in an active wireless network environment where devices continually communicate with a central station. These monitors occasionally leave the wireless coverage area during transport to a procedure room (for example, a trip to X-ray). When a monitor goes outside of the network range, monitoring continues locally.
76 Monitoring in communication with Acuity Welch Allyn Propaq LT Vital Signs Monitor Caution Do not leave a wireless monitor that is turned on outside of its wireless communication area for more than four hours. If your facility has unusual clinical workflows that require wireless devices to be off of the network for extended periods, do one or more of the following: • Extend the wireless coverage area to include areas where devices stay for longer periods of time.
Directions for Use Monitoring in communication with Acuity 77 Figure 79. Acuity patient list STEWART, ANN 7762940 Adult II 12:41:32 Rm 263 1mV/cm 80 HR/min 12 Resp/min NIBP mmHg SELECT PATIENT Enter new patient 4727937, BENET, CLAUDE 9542683, MERKER, HANS 7762940, STEWART, ANN 3398290, TAYLOR, MIKE 97 % SpO2 • If the patient’s name is already on the list, highlight the name and press .
78 Monitoring in communication with Acuity Press Welch Allyn Propaq LT Vital Signs Monitor to clear the message. When the monitor re-establishes communication with Acuity (after an interruption of at least 30 seconds), it prompts you to confirm that the patient being monitored is the same patient that was being monitored when communication was interrupted. Figure 82.
Directions for Use Monitoring in communication with Acuity 79 To disconnect from the network Disconnecting the monitor from the network while the monitor is outside of network range greatly extends battery life. However, connection with the network must first have been established before this is possible. Note 1. Highlight and press . 2. Highlight Disconnect, press , and confirm. Figure 84.
80 Monitoring in communication with Acuity Welch Allyn Propaq LT Vital Signs Monitor
81 5 Alarms and alerts Overview An Alarm warns of a patient condition, such as a vital-sign reading that is outside of acceptable limits. When an alarm occurs, the red light on the monitor flashes and the numerics of the violating alarm limits on the display turn red. If tones are not suspended, the alarm tone sounds. An Alert warns of an equipment condition, such as a low battery or a detached lead.
82 Alarms and alerts Welch Allyn Propaq LT Vital Signs Monitor Figure 86. Sample alarm screen HALL, ROBERT E. 3456187 Adult 3:00:06P Rm 239 II 1mV/cm 140/78 31 88 % 125 HR/min NIBP mmHg (102) Resp/min SpO2 @2:47P Manual Figure 87. Sample alert screen HALL, ROBERT E. 3456187 Audible Alarms Silenced Adult I 1mV/cm 80 HR/min 140/78 NIBP mmHg (102) @2:47P Manual 12:41:32 Rm 239 12 Resp/min 97 % SpO2 EQUIPMENT ALERT ECG Fault X LL Lead Failed ACKNOWLEDGE 2.
Directions for Use Alarms and alerts • 83 If it is configured to Always On and you set Suspend Audible Alarms to On, then the alarm tone remains suspended until: Note • you set Suspend Audible Alarms to Off or • monitor power is turned off and then turned on again or • the monitor is reconfigured Suspend Audible Alarms does not affect the behavior of the alarm silence/ reset feature ( ).
84 Alarms and alerts Welch Allyn Propaq LT Vital Signs Monitor • A countdown timer appears below the line to indicate the time remaining in the suspension period. • ‘Audio alarms suspended’ appears in yellow in the upper left corner of the screen. • When the suspension period elapses, the alarm tone is again enabled.
Directions for Use Alarms and alerts 85 About ParamSet Using ParamSet, you can quickly widen the alarm limits by a configured percentage (relative to the patient’s alarming reading) for any vital sign. For information about ParamSet, see “ParamSet Settings” on page 133. Responding to an alarm An alarm condition is indicated on the monitor in the following ways: • The RED indicator (rightmost of three) above the display screen flashes. • The numerics for the vital sign in alarm are displayed in RED.
86 Alarms and alerts Welch Allyn Propaq LT Vital Signs Monitor To respond at Acuity to an alert For a monitor in communication with Acuity, either the monitor or Acuity can detect alarms and alerts. The alarms are displayed in both places. See Acuity Directions For use. About battery charge status Table 6.
Directions for Use Alarms and alerts 87 Alert messages and status messages Table 7. Alert messages Alert type ECG NIBP Message Possible cause and suggested response ECG Fault. XX lead failed. Lead XX (LA, LL, RA, C, or RL) has very poor contact or no contact with the patient. Check for proper connection and replace the electrode if needed. ECG Fault. XX, XX leads failed. Leads XX and XX (any two leads on a 5-lead cable) have very poor contact or no contact with the patient.
88 Alarms and alerts Welch Allyn Propaq LT Vital Signs Monitor Table 7. Alert messages Alert type Network Communication Battery Message Possible cause and suggested response NIBP Fault. No valid blood pressure found. The patient mode setting is incorrect or the wrong hose or cuff is being used for the current patient mode. NIBP Fault. Calibrating. Please wait. The monitor periodically calibrates (zeroes) the NIBP channel to make sure it can properly make NIBP measurements.
Directions for Use Alarms and alerts 89 Table 7. Alert messages Alert type Message Charger SpO2 Masimo only Resp General Possible cause and suggested response Charger Fault. Service charger. Service required. Charger Disabled. Battery temperature too high or low. The battery is too cold or too hot to charge. Normalize the battery temperature before attempting to charge it. Battery Fault. Replace battery.
90 Alarms and alerts Welch Allyn Propaq LT Vital Signs Monitor Table 8. Status messages Message Notes NIBP calibrating. The monitor will resume operation when the calibration is complete. Button pad locked. Buttons (including the on/off button) remain locked until you unlock them or until an alarm or an alert sounds. Check SpO2 sensor. The sensor cable is faulty, incorrectly inserted, or incompatible. This message persists until a valid cable is connected or until monitor power is cycled.
91 6 Storing and reviewing patient data Overview The monitor stores up to 24 hours (at one-minute intervals) of trends information for the patient being monitored. When data storage is at capacity, the data from each new reading replaces the data from the oldest. Table 9. Store and review patient data All monitors Type of data Wireless monitor TRENDS • Tabular or graphical.
92 Storing and reviewing patient data Welch Allyn Propaq LT Vital Signs Monitor Reviewing data at the monitor Snapshots and trend data can be reviewed at the monitor. Reviewing snapshots To review snapshots From any main display screen (such as Large numerics, Dual waveform...): 1. Highlight 2. Press , HR/PR, SpO2, NIBP, Resp, or . . 3. Highlight Snapshots and press . The Snapshots display appears. Figure 92.
Directions for Use Storing and reviewing patient data 93 6. To change the source of the displayed waveform, see “To change the waveform selection” on page 55. 7. To change the size of the waveform, see “To change the waveform size” on page 56. 8. From this display, you can switch to another type of display or you can exit and return to the primary display. To do this, highlight Snapshots and press .
94 Storing and reviewing patient data Welch Allyn Propaq LT Vital Signs Monitor Figure 95. Trends display, 1-minute interval STEWART, ANN 7762940 Adult 120/72(88) 04:45:10P Rm 263 Tabular 59 14 98 % Time HR/min NIBP mmHg Resp/min SpO2 4:42P 4:41P 4:40P 4:40P 4:39P 4:39P 4:38P 4:37P 4:36P 4:35P 4:35P 4:34P 60 58 59 59 60 60 59 60 60 60 60 59 n/a n/a n/a n/a n/a 118/71(87) n/a n/a n/a n/a n/a 119/72(88) 14 13 14 14 13 14 14 14 13 14 13 13 99 98 98 99 98 98 98 99 98 98 99 98 4.
Directions for Use Storing and reviewing patient data Figure 98.
96 Storing and reviewing patient data Welch Allyn Propaq LT Vital Signs Monitor
97 7 Printing patient data Overview You can print patient data in two ways. • If the monitor is in communication with an Acuity Central Station, press vital signs at the Acuity system printer. to print • If a PC with the Autoprint Utility is connected to a cradle with the USB option, you can print directly to a printer connected to the PC simply by inserting the monitor in the cradle. This chapter tells you how to print with the Autoprint Utility.
98 Printing patient data Welch Allyn Propaq LT Vital Signs Monitor The printed data includes the following: In the Trends printout: • Up to 24 hours of HR/PR, RR/BR, and SpO2 trends (one set of measurements per row, with a 15-minute interval between rows). • SpO2 spot checks and NIBP measurements, interspersed among the rows of trends data. In the Snapshots printout: • Up to twenty snapshots (21-second intervals of full-disclosure patient data, including waveforms and numerics).
Directions for Use Printing patient data Note 99 For monitors equipped with Masimo SpO2 technology, the printout includes Perfusion Index readings. Figure 100. Printout (trends page) Monday, June 20, 2005 09:23:28 HALL, ROBERT E.
100 Printing patient data Welch Allyn Propaq LT Vital Signs Monitor Figure 101. Printout (snapshots page) Monday, June 20, 2005 11:44:55 HALL, ROBERT E. Name 3456187 Patient ID # 239 Location 1 of 1 Page Snapshot 25 mm/sec Adult (Extended) 11:44:55 HR = 58 RR = OFF SpO2 = OFF NIBP = OFF RR = 16 SpO2 = OFF NIBP = OFF RR = 15 SpO2 = OFF NIBP = OFF II 1.00mV 1 11:45:02 III 1.00mV V 1.00mV 1 11:45:02 HR = 60 II 1.00mV 2 11:45:09 III 1.00mV V 1.
Directions for Use Printing patient data 101 2. Click On or Off. You can use the AutoPrint Utility to print automatically (the default function) or manually. Printing manually To print manually 1. Start -> All Programs -> Welch Allyn -> Propaq LT Monitor -> AutoPrint On-Off 2. Click Off. 3.
102 Printing patient data • Welch Allyn Propaq LT Vital Signs Monitor www.adobe.com Welch Allyn does not recommend one PDF writer tool over another. 2. On the PC, change the default printer to the newly installed PDF writer, as follows: a. Open the Control Panel. b. Double-click Printers and Faxes. c. In the Printer and Faxes dialog, right-click on the newly installed PDF writer and check Set as Default. 3.
103 8 Monitor configuration Overview The monitor can be configured in various ways. Each facility can define as many default configurations as required to both meet the needs of individual departments and units and to conform to the facility’s general methods and standards. • Each time the patient mode is changed, the monitor settings return to the default configuration settings for that patient mode.
104 Monitor configuration Welch Allyn Propaq LT Vital Signs Monitor When a custom configuration is downloaded to the monitor, the settings and behaviors defined in the custom configuration become the power-up defaults for the monitor, and the factory configuration is no longer in effect. However, the factory configuration information can always be restored by qualified service personnel and can also be restored by downloading the configuration file PropaqLTFactoryConfig.mnt from a PC to the monitor.
Directions for Use Monitor configuration 105 Parameter configuration matrix This table lists the factory settings and the possible values for all parameters, and indicates which parameters and settings can be defined by Acuity, through the Configuration Utility, or by the clinician through the monitor buttons. Note This table shows only those parameters that can be changed.
106 Monitor configuration Welch Allyn Propaq LT Vital Signs Monitor Table 10. Parameter configuration matrix (continued) Parameter Factory setting Possible values Where set Time Format 24-Hr 12-Hr, 24-Hr Monitor, Configuration Utility Configured Language English Dutch, English, French, German, Italian, Polish, Portuguese, Spanish, Swedish Monitor, Acuity, Configuration Utility Decimal Format Dot Comma, Dot Configuration Utility Date Format MM/DD/YY MM/DD/YY, DD.MM.
Directions for Use Monitor configuration 107 Table 10. Parameter configuration matrix (continued) Parameter Factory setting Possible values Where set ECG aVF Wave Size 1 mV/cm 8 mV/cm, 4 mV/cm, 2 mV/cm, 1 mV/cm, Monitor 0.5 mV/cm, 0.2 mV/cm SpO2 Wave Size 2x 1x, 2x, 4x, 8x Monitor Resp Wave Size 2x 0.
108 Monitor configuration Welch Allyn Propaq LT Vital Signs Monitor Table 10.
Directions for Use Monitor configuration 109 Table 10.
110 Monitor configuration Welch Allyn Propaq LT Vital Signs Monitor Table 10. Parameter configuration matrix (continued) Parameter Factory setting Possible values Where set Auto Interval (minutes) 15 1, 2, 3, 5, 10, 15, 30, 60 Monitor, Configuration Utility Mode Manual Manual, Auto Monitor, Acuity Smartcuf Off On, Off Monitor, Acuity, Configuration Utility Systolic Upper Alarm Limit 145 mmHg 19.3 kPa 32 - 160 mmHg 4.3 - 21.3 kPa Monitor, Acuity Systolic Lower Alarm Limit 75 mmHg 10.
Directions for Use Monitor configuration 111 Table 10.
112 Monitor configuration Welch Allyn Propaq LT Vital Signs Monitor
113 9 PC utility Introduction Use the Propaq LT Monitor PC utility to install either or both of the following utilities on a PC: • Propaq LT Monitor Configuration utility • Propaq LT Monitor AutoPrint utility These utilities can then be used to control the behavior of any Propaq LT monitor. Propaq LT Monitor Configuration utility The monitor comes fully configured and ready for operation.
114 PC utility Welch Allyn Propaq LT Vital Signs Monitor Installation System requirements PC on which you have administrator privileges CPU Hard drive CD ROM drive Available RAM Available Port OS 800 MHz or faster 20 GB or more 256 MB or more USB 1.1 Windows 2000 or Windows XP Printer installed and configured Resolution 300 dpi minimum; 600 dpi or higher recommended Propaq LT Monitor PC Utility software (CD) Procedure Launching the installation 1.
Directions for Use PC utility 115 Figure 104. Accepting the terms of the software license 4. Determine whether you want to install the Configuration utility only, the AutoPrint utility only, or both utilities. Figure 105. Complete or Custom installation • To install both utilities, click Complete. • To install only one of the utilities, click Custom and, in the next screen, select the utility you want to install. Figure 106.
116 PC utility Welch Allyn Propaq LT Vital Signs Monitor Figure 107. Menu of installation options 5. The installation prompts you occasionally for information. At each prompt, provide the requested information and then click Next. The installation continues until it is finished. Using the Configuration Utility to configure the monitor Creating a configuration file 1. Start the Configuration utility. Double-click the Configuration icon. The Welcome screen appears. 2.
Directions for Use PC utility 117 Modifying a configuration file To modify a configuration file 1. Open the file. 2. Modify the settings. 3. Save the file. Configuring a monitor See “System requirements” on page 114. To configure a monitor 1. Connect the PC to a cradle, using the supplied USB cable. 2. Seat the monitor in the cradle. 3. If the monitor is off, turn it on and select Start New Patient. 4. Start the utility. 5. Select the configuration file that you want to upload. 6. Click Monitor > Send.
118 PC utility Welch Allyn Propaq LT Vital Signs Monitor Configuration settings Configuration pages Figure 108. Configuration file display: Identification page Click the page tabs (Identification, Default Settings, Display Settings, Mode Settings, ParamSet, and Feature Enable) to navigate to the settings of interest. Identification Name the configuration, define monitor ownership and support responsibility, and define parameters related to your locale. Each string is limited to 40 characters.
Directions for Use PC utility 119 Configuration Utility command options File New The utility opens a new configuration file with all settings at the factory values. Change the settings as needed and save the file under a new name. Open Open a configuration file. You can modify it (Save), use it as the starting point for a new file (Save As), or close it (Exit). Save Save the changes to the open configuration file. Save As Save the open configuration file under a new name.
120 PC utility Welch Allyn Propaq LT Vital Signs Monitor Message The configuration file did not install on the monitor. Verify all connections between monitor, cradle, and PC and try again. If problems continue, contact Welch Allyn customer service. The configuration File: [file name] that you are attempting to read is in the wrong format. Please select another file. The configuration file was modified outside of the Configuration Utility. Receive operation failed.
Directions for Use PC utility 121 The feature Back Light Lockout Enable has a factory-default value of Yes (enabled), so it is, by default, available to clinicians. To make this feature unavailable to clinicians, change the value to No in the custom configuration. Factory defaults remain in effect until you reconfigure the monitor by downloading a custom configuration file to it from the PC or until you replace them with the current settings.
122 PC utility Welch Allyn Propaq LT Vital Signs Monitor
Directions for Use 123 Configuration worksheet Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124 Define monitor ownership and support responsibility, define parameters related to your locale, and name the configuration. Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126 Define the power-up settings for monitor behavior. Display Settings . . . . . . . . .
124 Welch Allyn Propaq LT Vital Signs Monitor Identification Facility Facility Name (Up to 40 characters.) Contact Name (Up to 40 characters.) Department or Unit (Up to 40 characters.) Contact Telephone (Up to 40 characters.) Country Settings Language __ English __ Dutch __ French __ German __ Italian __ Polish __ Portuguese __ Spanish __ Swedish __ Japanese Date Format __ mm/dd/yy __ dd.mm.
Directions for Use 125 Time Format __ 12-hour AM/PM __ 24-hour Decimal Format __ Dot __ Comma You can configure the monitor to use a dot (for example, 195.18) or a comma (195,18) to display the decimal point. Regulatory Set __ US __ European EC __ Japan Metrology Configuration Configuration Name Configuration Name (Up to 40 characters.
126 Welch Allyn Propaq LT Vital Signs Monitor Default Settings Default Selections Patient Mode Default __ Adult __ Pediatric __Neonatal Alarm Tone Level Default __ Low __ Medium __ High ECG Bandwidth Default __ Monitor __ Extended ECG Bandwidth refers to the frequency range used to display and print patient data. • Monitor bandwidth is less susceptible to artifact such as patient motion. • Extended bandwidth, although more susceptible to artifact, can provide for more detailed data analysis.
Directions for Use 127 Audible Alarms Suspension Time Default __ Disable (alarm tones cannot be suspended) __ 90 seconds __ 2 minutes __ 3 minutes __ 4 minutes __ 5 minutes __ 10 minutes __ 15 minutes __ 30 minutes __ 60 minutes __ Always On (alarm tones always suspended) While monitoring a patient, the clinician can use the settings menu to suspend all alarm tones for all parameters for a configurable period.
128 Welch Allyn Propaq LT Vital Signs Monitor Display Settings The monitor interface includes a Display button: . Pressing quickly through the three views you specify below. cycles the display In the configuration, you can choose any three of the five display choices shown below for quick display with .
Directions for Use 129 Display Format Display 1 Display 2 Display 3 __ Large Numerics __ Large Numerics __ Large Numerics __ Single Waveform __ Single Waveform __ Single Waveform __ Dual Waveform __ Dual Waveform __ Dual Waveform __ Tabular Trends __ Tabular Trends __ Tabular Trends __ Tabular Trends with Waveform __ Tabular Trends with Waveform __ Tabular Trends with Waveform __ None __ None Display 1 Display 2 Display 3 __ Lead I __ Lead I __ Lead I __ Lead II __ Lead II __ L
130 Welch Allyn Propaq LT Vital Signs Monitor Display Format (continued) If Dual Waveform display is enabled, display this signal as the default lower waveform: Display 1 Display 2 Display 3 __ Lead I __ Lead I __ Lead I __ Lead II __ Lead II __ Lead II __ Lead III __ Lead III __ Lead III __ Lead V __ Lead V __ Lead V __ aVR __ aVR __ aVR __ aVL __ aVL __ aVL __ aVF __ aVF __ aVF __ SpO2 __ SpO2 __ SpO2 __ Resp __ Resp __ Resp If Tabular Trends display is enabled, this is the
Directions for Use 131 Mode Settings Default Resp Lead Adult Pediatric Neonatal __ Lead 1 (RA-LA) __ Lead 1 (RA-LA) __ Lead 1 (RA-LA) __ Lead 2 (RA-LL) __ Lead 2 (RA-LL) __ Lead 2 (RA-LL) NIBP Mean Numerics Adult Pediatric Neonatal Small Small Small Large Large Large Off Off Off Note Selecting Large reduces the size of the systolic and diastolic displays.
132 Welch Allyn Propaq LT Vital Signs Monitor Resp On Default Adult Pediatric Neonatal (Yes) (Yes) (Yes) (No) (No) (No) Alarm Limits HR/PR Adult Pediatric Neonatal Upper On Off 120 ___ Upper On Off 150 ___ Upper On Off 200 ___ Lower On Off Lower On Off Lower On Off 100 ___ 50 ___ 50 ___ Resp Adult Pediatric Neonatal Upper On Off 30 ___ Upper On Off 45 ___ Upper On Off 60 ___ Lower On Off 5 ___ Lower On Off 10 ___ Lower On Off 10 ___ SpO2 Adult Pediatric Neonatal U
Directions for Use 133 ParamSet Settings ParamSet Enable (Yes) (No) Note If you select HR/PR ParamSet % Resp ParamSet % SpO2 ParamSet % (No), disregard the rest of the ParamSet settings.
134 Welch Allyn Propaq LT Vital Signs Monitor Note The ParamSet default values can be changed only by downloading another monitor configuration. When monitor power is cycled, the factory default alarm limits and ParamSet default values are again in effect. ParamSet can be used multiple times for any alarm limit. Each time you press ParamSet for a given vital sign, the alarm limit expands by the configured percentage.
Directions for Use 135 Feature Enable System Features Enable Wireless Communications (with Acuity) (Yes) (No) For wireless monitors that must communicate with Acuity, this must be set to this is set to (No), the monitor cannot communicate with Acuity. (Yes). If For standalone monitors that do not communicate with Acuity, always set this to (No). If this is set to (Yes), the configuration file is not valid and cannot be uploaded to the monitor.
136 Welch Allyn Propaq LT Vital Signs Monitor Back Light Lockout Enable (Yes) (No) If back light lockout is enabled, the 3-button combination locks out the monitor display back light until the combination is repeated or until an alarm or alert condition arises. Buttons Lockout Enable (Yes) (No) If you enable buttons lockout, the 3-button combination locks out normal button operation until the combination is repeated or until an alarm or alert condition arises.
Directions for Use 137 Masimo SpO2 Perfusion Index (Enable) (Disable) Signal IQ (Enable) (Disable) Sensitivity (Enable) (Disable) If enabled: Normal, Maximum, APOD FastSAT (Enable) (Disable) Default On (Yes) (No) Averaging Time (Enable) (Disable) If enabled: 2, 4, 8, 10, 12, 14, 16 (seconds)
138 Welch Allyn Propaq LT Vital Signs Monitor Authorization The information on this page relates only to this worksheet, and is not entered into the monitor configuration. Users are encouraged to provide their own procedures for defining, categorizing, storing, and approving configuration files.
139 10 Maintenance Recharging the battery The monitor battery is recharged whenever the monitor is placed into a cradle that is connected to a suitable power source. Recharging occurs whether the monitor is on or off. While in a cradle that is connected to AC power, the monitor also receives operating power from the cradle, thus preserving battery power for use when the monitor is removed. To recharge the battery 1. Confirm that the green “power” light (the right of two on the cradle) is on.
140 Maintenance Welch Allyn Propaq LT Vital Signs Monitor WARNING Do not incinerate, submerge, crush, disassemble, or autoclave the lithium-ion battery. Caution Do not remove or replace the battery unless you are a Welch Allyn qualified service technician. Caution Do not use the monitor while the battery is being replaced. Inspecting and cleaning the monitor and accessories WARNING Do not autoclave the monitor, the cradle, the AC power adapter, or any accessories.
Directions for Use Maintenance 141 Caution Never use any of the following solutions or similar products to clean the equipment: butyl alcohol, denatured ethanol, Freon™, mild chlorine bleach solution, isopropyl alcohol (except for the SpO2 cable), trichloroethane, trichloroethylene, acetone, Vesphene II, Enviroquat®, Staphene®, Misty®, Glutaraldehyde, Formula 409®, or Fantastik®. Table 11.
142 Maintenance Welch Allyn Propaq LT Vital Signs Monitor
143 A Specifications ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143 Resp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146 SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147 NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
144 Specifications Welch Allyn Propaq LT Vital Signs Monitor Lead-fail sense current 70 nA DC typical for active leads 140-280 nA DC typical for reference electrode, depending on number attached. Tall T-wave rejection Meets AAMI (USA) EC13-2002, section 4.1.2.1.c, up through 1.
Directions for Use Specifications Pacer pulse rejection Does not count as heartbeats approximately 95% of pacemaker pulses within pacer pulse detection range*, with or without AAMI (EC 13-2002) tails of 4, 8, 15, 25, 50, 75, or 100 ms decay time constant, whose tail amplitudes meet either EC 13’s method A or method B criteria for single pacer pulses, or A-V sequential pulses (150 ms and 250 ms separation), all per AAMI tests 4.1.4.1 and 4.1.4.2.
146 Specifications Welch Allyn Propaq LT Vital Signs Monitor Resp Characteristic Specification Connector (shared with ECG) Hypertronics 9-pin, female D01 latching connector Input protection Electrosurgery and defibrillator protected and Resp fully functional when used with any of the ECG cables specified in the accessory list or at www.welchallyn.com. Electrosurgery interference suppression Yes, but not quantifiable.
Directions for Use Specifications 147 SpO2 All sensors have a measurement range of 70% - 100%. SpO2 complies with EN ISO 9919:2005. Characteristic Specification Saturation (% SpO2) Range Resolution 1% to 100%, but limited by sensors and by sensor manufacturers’ technology 1% Alarm limits 50% to 98% (lower) 52% to 100% (upper) Probe accuracy (adult, pediatric, neonate) Accuracy for saturation levels below 70% is unspecified.
148 Specifications Welch Allyn Propaq LT Vital Signs Monitor NIBP Characteristic Specification Method Oscillometric Regulatory standards Meets AAMI SP10:2002, IEC 60601-1:1990, IEC 60601-2-30:1999(E); EN 1060-1:1996, EN 1060-3:1997 Control Automatic (at selected intervals), turbo, and manual Auto intervals 1, 2, 3, 5, 10, 15, 30, and 60 minutes Smartcuf Available when ECG is used Reported pressures Systolic, diastolic, and mean plus real-time manometer pressure Pressure resolution 1 mmHg
Directions for Use Specifications Normal overpressure limit Adult Pediatric Neonate 280 mmHg 200 mmHg 141 mmHg Single fault overpressure limit Adult Pediatric Neonate 308 mmHg 220 mmHg 154 mmHg Leak rate (manufacturing spec) After a 1-minute settling period, not more than 1 mmHg per second at 200 mmHg when connected to a volume of at least 15 cc.
150 Specifications Welch Allyn Propaq LT Vital Signs Monitor Patient alarm high-priority limit violation [100 ms on, 80 ms off, 100 ms on, 80 ms off, 100 ms on, 260 ms off, 100 ms on, 80 ms off, 100 ms on, 1 sec off, 100 ms on, 80 ms off, 100 ms on, 80 ms off, 100 ms on, 260 ms off, 100 ms on, 80 ms off, 100 ms on, 4 sec off]; repeat. Equipment alert medium priority [160 ms on, 200 ms off, 160 ms on, 200 ms off, 160 ms on, 3.96 sec off]; repeat.
Directions for Use Specifications SpO2 Adult Pediatric Neonate 100% 100% 95% Adult Pediatric Neonate 90% 90% 85% NIBP Systolic Adult Pediatric Neonate 220 145 100 Adult Pediatric Neonate 75 75 50 NIBP Diastolic Adult Pediatric Neonate 110 100 70 Adult Pediatric Neonate 35 35 30 NIBP MAP Adult Pediatric Neonate 120 110 80 Adult Pediatric Neonate 50 50 35 151 Display Characteristic Specification Type Color transflective; LCD module Resolution 320 x 240 pixels Quarter VGA Active vi
152 Specifications Welch Allyn Propaq LT Vital Signs Monitor WARNING In the event a radar signature is detected, the access point moves to a new channel, which can temporarily interrupt patient monitoring. Caution If the device is operated near a military radar, the radar could cause damage to the device. Caution Some countries restrict the use of 5-GHz bandwidths. The 802.11a radio in the Model 802LTA monitors uses only the channels indicated by the access point with which the radio associates.
Directions for Use Specifications 153 Flat-panel display for the Large Color Display Interface These specifications are based on a 17-inch (diagonal measure) display with a 5:4 aspect ratio or a 22-inch (diagonal measure) display with a 16:9 aspect ratio. Size 17” or 22“ diagonal Resolution Must be compatible to 1024 dots x 768 lines. 49.72KHz Horizontal, 62.46Hz vertical and 66.818 MHz Pixel Clock. Video signal Analog RGB (0.
154 Specifications Welch Allyn Propaq LT Vital Signs Monitor Environmental Caution The monitor might not meet performance specifications if it is not used or stored within these environmental specifications. Note The specifications apply when the battery is installed and the battery cover is in place. Characteristic Specification Operating temperature 0° to 40° C Shipping and storage temperature -20° to 60° C for monitor and cradle.
Directions for Use Specifications Degree of protection against electric shock, for parts applied to patients See monitor labels. Type CF, defibrillator protected. Method of cleaning Not suitable for autoclaving. See “Inspecting and cleaning the monitor and accessories” on page 140. Flammable anesthetics Not suitable for use with flammable anesthetics. Physical dimensions Monitor height 5.4 in (13.71 cm) Monitor width 7.5 in (19.05 cm) Monitor depth 2.1 in (5.
156 Specifications Welch Allyn Propaq LT Vital Signs Monitor Power Characteristic Specifications Operating times on a new fully-charged battery for ambient temperature 20º C Standalone Continuous monitoring with minimal display use (defined as 5 minutes display use per hour) Acuity connected (Models 802LTAN, 802LTAS, 802LTRN and 802LTRS) In communication, with continuous monitoring and minimal display use (5 min display use/hour) ECG and Resp only: 24 hrs ECG, Resp, SpO2. NIBP every 15 min.
Directions for Use Specifications 157 Battery Characteristic Specification Type 7.4 V, rechargeable, lithium ion. Field-replaceable by service technician. Capacity > 1800 mA-hr minimum Charger Contained in the cradle. Batteries are charged in place in the monitor. Fuse rating 5A, 125V – User cannot access the fuse. Note: Internal electronic overload circuitry is used as the primary method of protection. This circuit resets itself when an overload is removed.
158 Specifications Welch Allyn Propaq LT Vital Signs Monitor Vibration, random 0.02g2/Hz from 10 to 500 Hz, ramping down to 0.002g2/Hz at 2000 Hz. Superimposed sine frequencies of 30 Hz at 2.5 g and 60, 90, and 120 Hz at 1.5 g. Operating 1 hour per axis, 3 hours per test. Electromagnetic compatibility (EMC) with USB option EN60601-1-2:2001 Class A without USB option EN60601-1-2:2001 Class B Physical Depth 6.0 in (15.24 cm) Width 7.6 in (19.3 cm) Height 5.7 in (14.
159 B Compliance General This 802.11a Wireless PC Card must be installed and used in strict accordance with the manufacturer’s instructions as described in the user documentation that comes with the product. This product contains encryption. It is unlawful to export out of the U.S. without obtaining a U.S. Export License. Federal Communications Commission (FCC) This device complies with part 15 of the FCC Rules.
160 Compliance Welch Allyn Propaq LT Vital Signs Monitor The correction of interference caused by such unauthorized modification, substitution or attachment will be the responsibility of the user. Industry Canada (IC) emissions This device complies with RSS 210 of Industry Canada. Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of this device.
Directions for Use Compliance Portuguese Welch Allyn declara que este RLAN device está conforme com os requisitos essenciais e outras disposições da Directiva 1999/5/CE. Slovak Welch Allyn týmto vyhlasuje, ze RLAN device spĺňa základné po_iadavky a všetky príslušné ustanovenia Smernice 1999/5/ES. Slovene Šiuo Welch Allyn deklaruoja, kad šis RLAN device atitinka esminius reikalavimus ir kitas 1999/5/EB Direktyvos nuostatas.
162 Compliance Welch Allyn Propaq LT Vital Signs Monitor EMC Special precautions concerning electromagnetic compatibility (EMC) must be taken for all medical electrical equipment. • All medical electrical equipment must be installed and put into service in accordance with the EMC information provided in this document. • Portable and mobile RF communications equipment can affect the behavior of medical electrical equipment.
Directions for Use Compliance 163 Monitor, battery-operated The EMC specifications listed on pages 163 to 174 apply to Propaq LT monitors using any ECG cable listed in the accessory list or at www.welchallyn.com. For other EMC specifications, see “Monitor, battery-operated, with cable 008-0799-00/01” on page 175. Guidance and manufacturer’s declaration—electromagnetic emissions The battery-operated Propaq LT Series monitor is intended for use in the electromagnetic environment specified below.
164 Compliance Welch Allyn Propaq LT Vital Signs Monitor Guidance and manufacturer’s declaration—electromagnetic immunity The battery-operated Propaq LT Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
Directions for Use Compliance 165 Guidance and manufacturer’s declaration—electromagnetic immunity The battery-operated Propaq LT Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
166 Compliance Welch Allyn Propaq LT Vital Signs Monitor Recommended separation distances between portable and mobile RF communications equipment and the battery-operated Propaq LT Series monitor The battery-operated Propaq LT Series monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
Directions for Use Compliance 167 Charging/communication cradle with monitor Guidance and manufacturer’s declaration—electromagnetic emissions The Charging/Communication Cradle with Propaq LT Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Charging/Communication Cradle with Propaq LT Series monitor should assure that it is used in such an environment.
168 Compliance Welch Allyn Propaq LT Vital Signs Monitor Guidance and manufacturer’s declaration—electromagnetic immunity The Charging/Communication Cradle with Propaq LT Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
Directions for Use Compliance 169 Guidance and manufacturer’s declaration—electromagnetic immunity The Charging/Communication Cradle with Propaq LT Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
170 Compliance Welch Allyn Propaq LT Vital Signs Monitor Recommended separation distances between portable and mobile RF communications equipment and the Charging/Communication Cradle with Propaq LT Series monitor The Charging/Communication Cradle with Propaq LT Series monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
Directions for Use Compliance 171 Monitor and cradle with large color display interface Guidance and manufacturer’s declaration—electromagnetic emissions The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface is intended for use in the electromagnetic environment specified below. The customer or the user of the The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface should assure that it is used in such an environment.
172 Compliance Welch Allyn Propaq LT Vital Signs Monitor Guidance and manufacturer’s declaration—electromagnetic immunity The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface is intended for use in the electromagnetic environment specified below. The customer or the user of the The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface should assure that it is used in such an environment.
Directions for Use Compliance 173 Guidance and manufacturer’s declaration—electromagnetic immunity The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface is intended for use in the electromagnetic environment specified below. The customer or the user of the Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface should assure that it is used in such an environment.
174 Compliance Welch Allyn Propaq LT Vital Signs Monitor Recommended separation distances between portable and mobile RF communications equipment and the Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
Directions for Use Compliance 175 Monitor, battery-operated, with cable 008-0799-00/01 The EMC specifications listed on pages 175 to 186 apply only to Propaq LT monitors using ECG cables 008-0799-00 and 008-0799-01. For general EMC specifications, see “Monitor, battery-operated” on page 163. Guidance and manufacturer’s declaration—electromagnetic emissions The battery-operated Propaq LT Series monitor is intended for use in the electromagnetic environment specified below.
176 Compliance Welch Allyn Propaq LT Vital Signs Monitor Guidance and manufacturer’s declaration—electromagnetic immunity The battery-operated Propaq LT Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
Directions for Use Compliance 177 Guidance and manufacturer’s declaration—electromagnetic immunity The battery-operated Propaq LT Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
178 Compliance Welch Allyn Propaq LT Vital Signs Monitor Recommended separation distances between portable and mobile RF communications equipment and the battery-operated Propaq LT Series monitor The battery-operated Propaq LT Series monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
Directions for Use Compliance 179 Charging/communication cradle with monitor, with cable 008-0799-00/01 Guidance and manufacturer’s declaration—electromagnetic emissions The Charging/Communication Cradle with Propaq LT Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Charging/Communication Cradle with Propaq LT Series monitor should assure that it is used in such an environment.
180 Compliance Welch Allyn Propaq LT Vital Signs Monitor Guidance and manufacturer’s declaration—electromagnetic immunity The Charging/Communication Cradle with Propaq LT Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
Directions for Use Compliance 181 Guidance and manufacturer’s declaration—electromagnetic immunity The Charging/Communication Cradle with Propaq LT Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
182 Compliance Welch Allyn Propaq LT Vital Signs Monitor Recommended separation distances between portable and mobile RF communications equipment and the Charging/Communication Cradle with Propaq LT Series monitor The Charging/Communication Cradle with Propaq LT Series monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
Directions for Use Compliance 183 Monitor and cradle with large color display interface, with cable 008-0799-00/01 Guidance and manufacturer’s declaration—electromagnetic emissions The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface is intended for use in the electromagnetic environment specified below.
184 Compliance Welch Allyn Propaq LT Vital Signs Monitor Guidance and manufacturer’s declaration—electromagnetic immunity The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface is intended for use in the electromagnetic environment specified below. The customer or the user of the The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface should assure that it is used in such an environment.
Directions for Use Compliance 185 Guidance and manufacturer’s declaration—electromagnetic immunity The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface is intended for use in the electromagnetic environment specified below. The customer or the user of the Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface should assure that it is used in such an environment.
186 Compliance Welch Allyn Propaq LT Vital Signs Monitor Recommended separation distances between portable and mobile RF communications equipment and the Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
187 Limited warranty This product is sold by Welch Allyn under the warranties set forth in the following paragraphs. These warranties are extended only to the end user with respect to the original purchase of this product directly from Welch Allyn or from Welch Allyn’s authorized distributors.
188 Limited warranty Welch Allyn Propaq LT Vital Signs Monitor
189 Index A accessories, 11 action, 31 Acuity communication, 44, 73 disconnecting, 79 indicators on the monitor display, 17 marking snapshot events at, 150 monitor data stored at, 91 overriding monitor settings, 103 patients, arrhythmia-prone (warning), 5 reconnecting a patient, 79 responding to alarms at, 86 silencing alarms at, 82 values imposed on monitor settings, 104 adult term defined, 49 alarm defined, 81 responding, 86 alarm limits changing quickly with ParamSet, 85 customizing, 84 disabled, 149 ra
190 Index dual waveform display, 19 E ECG changing waveform scale, 56 lead placement, 55 electrosurgery (warning), 4 equipment alert messages, 86 error detection, 44 F FastSAT, 64 FlexNet network, 1 H hazards general warnings, 4 heart rate (HR) vs.
Directions for Use pulse amplitude, 18 and NIBP measurement accuracy, 68 pulse rate effect on SpO2 readings, 60 pulse rate (PR) vs.
192 Index Welch Allyn Propaq LT Vital Signs Monitor Printed manual material number: 721483