ProBP 3400 tab
| OpenTopic | TOC Contents ProBP 3400 tab ...............................................................................................................................................3 Managed devices ............................................................................................................................................4 ProBP3400 error codes ...................................................................................................................................
OpenTopic | ProBP 3400 tab | 3 ProBP 3400 tab The ProBP 3400 tab is the single access point to all device information and actions that you may perform on a ProBP 3400 (except product license installations – see Install a license for details. It is divided into two main areas: a general device information pane and a secondary tabbed pane. General device information pane The general device information pane shows device identification, location information, and communication addresses.
| OpenTopic | Managed devices Managed devices Welch Allyn Service Tool provides features that can help you manage and maintain your inventory of supported Welch Allyn products. The Welch Allyn Service Tool install connects directly to Welch Allyn Customer Service. This system automatically receives firmware updates and feature upgrades for your supported products. Maintained device list Managing your product inventory begins with establishing your offline device list.
OpenTopic | ProBP3400 error codes | 5 ProBP3400 error codes The following tables list the error messages associated with the Welch Allyn ProBP 3400. Error code Related System Additional details 000000001 Persistent storage error EERPOM is corrupt or invalid 000000001 BP module failure Mod F NIBP board failure Error condition Definition / Description Required action Low battery 5 minutes remaining. The battery is extremely low.
| OpenTopic | ProBP3400 error codes Error condition Definition / Description Required action Set date and time. Either date or time not set at all or data has not been set. Set date and time. Ambient temperature outside Device outside operating temperature An alarm will sound, but you may operating range. Retry measurement. range. continue to take measurements. Maximum number of patient records saved. Oldest record overwritten.
OpenTopic | ProBP3400 error codes | 7 Error condition Definition / Description Required action Device is operating in battery mode. AC power cord has been disconnected. Click the OK button to dismiss the message. Plug the device into the outlet to dismiss the message. Powering down. Device is performing normal shutdown activities. System error; disconnect patient cables, power down, and restart. System failure: fault is detected that could adversely affect the product. Data did not transfer.
| OpenTopic | ProBP 3400 calibration and verification setup ProBP 3400 calibration and verification setup Test equipment part numbers The equipment detailed in the table below are used for ProBP 3400 calibration and verification. Contact Welch Allyn Customer Service to purchase or locate specific parts. Part number Description Quantity BP Y-tube, no fittings, 1/8 tube 1 4500-925 USB 2.
OpenTopic | ProBP 3400 calibration and verification setup | 9 1. Connect a USB cable to the service test box; connect the opposite end to a PC. 2. Connect a mini B USB cable to the ProBP 3400 and the opposite end to the service test box. 3. Connect the blood pressure hose to the NIBP module and the test volume. • • • • Connect a piece of tubing to the squeeze bulb with bleeding screw; connect the opposite end to the square fitting connector.
| OpenTopic | Calibrate and verify the device Calibrate and verify the device 1. Click within the device row, then click Perform All to activate the test sequence for the module. 2. Enter information about any replacement parts used during service of the product. After entering the part information, click Begin to start the tests. 3. Follow the onscreen prompts in the right side pane of the test window to progress through the test sequence.
OpenTopic | Configure tab | 11 Configure tab The configure tab enables all users the option to change the location ID and the asset tag number. Additionally, users may configure the ProBP 3400 to: • • • • • Enable or disable the device to emulate Spot Vitals Signs LXi. Set the time limit before entering standby mode due to inactivity. Set the time limit before powering off due to inactivity. Enable or disable serial emulation (Bluetooth default setting).