User Manual User guide

1
©2014 Welch Allyn, Inc.
MM 721358 Ver. C
Standard
laryngoscope blade assemblies
Directions for use
Intended
use
The
rigid laryngoscope is a device used to examine
and visualize
a patient’s airway
and
aid
placement of a tracheal
tube.
About this
document
This directions for use applies to Welch Allyn reusable standard laryngoscope blade
assemblies MacIntosh
REF:
6904X, English MacIntosh
REF:
6924X, and Miller REF:
6804X and REF 68470. Welch Allyn
reusable standard
laryngoscope blade assemblies may be
used
with
Welch Allyn standard laryngoscope handles 60200, 60300, 60400, 60305, and 60710.
Warnings and
cautions
WARNING: R
eusable
fiber optic
laryngoscope blade
assemblies must
be reprocessed
after
each use.
WARNING: The reprocessing procedure and the equipment and materials described must
be followed and conducted by persons trained and familiar
with
medical device
reprocessing.
WARNING: Consult cleaning and disinfecting agent manufacturer
instructions
for their
proper preparation and use.
WARNING: Repeated
reprocessing may
degrade
elements of the device.
Follow
inspection
procedures to assure damage has not occurred to the blade assembly.
WARNING: Only trained personnel shall use a laryngoscope for
intubation.
WARNING: Do not reprocess laryngoscope blade assemblies used on
patients
with
Creutzfeldt-Jakob disease. These instruments should be discarded.
WARNING: Discard any component that shows evidence of damage or deterioration.
WARNING: Do not modify this equipment. Any modification of this
equipment
can lead to
patient injury. Any modification of this equipment voids the
product warranty.
WARNING: Personnel
shall follow their facility policies and procedures and
wear
appropriate
personal protective equipment when handling
potentially
contaminated
equipment.
WARNING:
Laryngoscope equipment is not suitable for use in intense
magnetic
fields.
WARNING: Lamps if left illuminated, could generate sufficient heat to cause burns.
CAUTION: Only use lamps specified. Failure
to follow these instructions may cause damage
or poor performance of this device.
CAUTION:
Only use neutral pH
(6.0-8.0)
enzymatic type cleaners.
CAUTION:
Only use neutral pH
(6.0-8.0)
ortho-phthalahyde
(OPA)
or glutaraldehyde
disinfectants.
CAUTION:
Do not use other cleaners or
disinfectants.
CAUTION:
Do not remove lamp during reprocessing. Damage to device could occur.
CAUTION:
REF 68470 is not compatible with any sterilization process.