5.1. Manufacturers Declaration - Electromagnetic Emission
The FireCR system is intended for use in the electromagnetic environment specified below. The
customer or the user of FireCR system should assure that it is used in such an environment
Emission test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The FireCR system uses RF energy only for its
internal function. Therefore. Its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment
RF emissions
CISPR 11
Class B
The FireCR system is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supplies buildings used for domestic
purposes.
Harmonics emission
IEC 61000-3-2
A
Voltage fluctuation
IEC 61000-3-3
Complies
5.2. Manufacturers Declaration - Electromagnetic Immunity
The FireCR system is intended for use in the electromagnetic environment specified below. The
customer or the user of FireCR system should assure that it is used in such an environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
Environment -guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV Contact
8 kV Air
6 kV Contact
8 kV Air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%
Electrical fast
Transient / burst
IEC 61000-4-4
2kV for power supply
lines 1kV for input/output
lines
2kV for power supply
lines
1kV for input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.