FireCR
Power frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
3.0 A/m 3.0 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short
Interruptions and
Voltage variations
on power supply
input lines
IEC 61000-4-11
<5% Uт (>95% dip in Uт)
for 0.5cycle
40% Uт (60% dip in Uт )
for 5 cycle
70% Uт (30% dip in Uт)
for 25 cycle
<5% Uт (<95% dip in Uт )
for 5 s
<5% Uт (>95% dip in Uт)
for 0.5cycle
40% Uт (60% dip in Uт )
for 5 cycle
70% Uт (30% dip in Uт)
for 25 cycle
<5% Uт (<95% dip in Uт )
for 5 s
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the BSVD-1000
system requires continued
operation during power mains
interruptions, it is
recommended that the BSVD-
1000 system be powered from
an uninterruptible power
supply or a battery
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
Portable and mobile RF
communications equipment
should be used no closer to
any part of the FireCR system,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
Radiated RF
IEC 61000-4-3
3 V/m
80.0 MHz to 2.5 GHz
3 V/m
80.0 MHz to 2.5 GHz
Recommended separation
distance
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as deter-mined
by an electromagnetic site
survey,
(a) Should be less than the
compliance level in each
frequency range (b).
Interference may occur in the
vicinity of
equipment marked with the
following symbol: