5.3. Guidance and Manufacturer s Declaration
Electromagnetic Immunity
The FireCR Dental system
is intended for use in the electromagnetic environment specified
below. The customer or the user of FireCR Dental system
should assure that it is used in such an
environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
environment
-
guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80MHz
3 Vrms
150 kHz to 80 MHz
FireCR Dental system must be used
only in a shielded location with a
minimum RF shielding effectiveness
and, for each cable that enters the
shielded location with a minimum RF
shielding effectiveness and, for each
cable that enters the shielded location
Radiated RF
IEC 61000-4-3
3 V/m
80.0 MHz to 2.5GHz
3 V/m
80.0 MHz to 2.5GHz
Field strengths outside the shielded
location from fixed RF transmitters, as
determined by an electromagnetic site
survey, should be less than 3V/m.a
Interference may occur in the vicinity of
equipment marked with the following
symbol:
N
ote
1)
These guidelines may not apply
in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Note 2)
It is essential that the actual shielding effectiveness and filter attenuation of the shielded
location be verified to assure that they meet the minimum specification.
a
-
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field
strength outside the shielded location in which the EUT is used exceeds 3V/m, th
e EUT should be
observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as relocating
the EUT or using a shielded location with a higher RF shielding effectiveness and filter attenuation.