User's Manual

User Manual
FireCR Spark TM-701-EN
25
5.2. Manufacturers Declaration - Electromagnetic Immunity
The FireCR Spark system is intended for use in the electromagnetic environment specified
below. The customer or the user of FireCR Spark system should assure that it is used in such
an environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
Environment -guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV Contact
8 kV Air
6 kV Contact
8 kV Air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%
Electrical fast
Transient / burst
IEC 61000-4-4
2kV for power supply
lines 1kV for input/output
lines
2kV for power supply
lines
1kV for input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
3.0 A/m 3.0 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short
Interruptions and
Voltage variations
on power supply
input lines
IEC 61000-4-11
<5% Uт (>95% dip in Uт)
for 0.5cycle
40% Uт (60% dip in Uт )
for 5 cycle
70% Uт (30% dip in Uт)
for 25 cycle
<5% Uт (<95% dip in Uт )
for 5 s
<5% Uт (>95% dip in Uт)
for 0.5cycle
40% Uт (60% dip in Uт )
for 5 cycle
70% Uт (30% dip in Uт)
for 25 cycle
<5% Uт (<95% dip in Uт )
for 5 s
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the FireCR
Spark
system requires continued
operation during power mains
interruptions, it is
recommended that the FireCR
Spark
system be powered
from an uninterruptible power
supply or a battery
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
Portable and mobile RF
communications equipment
should be used no closer to
any part of the FireCR Spark
system, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance