5.4. Laser Safety Statement
The Computed Radiography Reader is Certified in the U.S. to Conform to the Requirements of
DHHS 21 CFR, chapter 1 Subchapter J for Class I(1) Laser Products, and Elsewhere is
Certified as a Class I(1) Laser Product Conforming to the Requirements of IEC 60825-1 :
2007. Class I(1) Laser Products are not Considered to be Hazardous. The Laser System and
Computed Radiography Reader are Designed so there is never any Human Access to Laser
Radiation above a Class I(1) level during normal Operation, user Maintenance or Prescribed
Service Condition.
• Wavelength : 658 nm (Typ.)
• Beam Divergence
- Paraller : 9.5 degrees (-2.5/+2.5)
- Perpendicular : 17 degrees (-3/+3)
• Maximum Power of Energy Output : 80 mW (CW)
WARNING
Never operate or service the product with the protective cover removed
from Laser/Reader assembly.
The reflected beam, although invisible, can damage your eyes.
When using this product, these basic safety precautions should always be
followed to reduce risk of fire, electric shock and personal injury
CAUTION
Use of controls or adjustments or performance of procedures other than
those specified herein may result in hazardous radiation exposure