power (W) of transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz
800 MHz to 2.5
GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.70 3.70 7.37
100 11.70 11.70 23.30
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
Immunity and Compliance Level
Immunity test IEC 60601 Test Level Actual Immunity Level Compliance Level
Conducted RF
IEC 61000-4-6
3 Vrms, 150 kHz to 80
MHz
3 Vrms, 150 kHz to 80
MHz
3 Vrms, 150 kHz to 80
MHz
Radiated RF
IEC 61000-4-3
3 V/m, 80 MHz to 2.5
GHz
3 V/m, 80 MHz to 2.5
GHz
3 V/m, 80 MHz to 2.5
GHz