User's Manual
Table Of Contents
- Introduction and Basics
- Introduction
- Safety Information
- Installing and Moving Sprite
- Safety
- External Machine Diagram
- Machine Specifications
- Contact Information
- Basic Operation
- Sample Placement
- Contact Information
- Components
- Main Screen
- Begin a New Scan Screen
- Scanning Screen
- Scan Review Screen
- Processing Setup Screen
- Results Screen
- _
- Locate a Scan Screen
- Background Processing Screen
- Utilities Screen
- System Sleep Button
- Touch screen Calibration Button
- System Shutdown Button
- Storage Device Management Button
- Network Configuration Button
- One Touch Scan Setup Button
- Submit Trouble Ticket Button
- Reset Processing Engine Button
- Reset System Board Service Button
- Reset X-ray Service Button
- Reset Scan Engine Button
- System Restart Button
- System Status Screen
- X-ray System Maintenance Screen
- Common Elements
- Technical Support
Sprite User's Manual
58
X-ray System Maintenance Screen
This screen allows you to manually begin or schedule automatic maintenance procedures. These
procedures – X-ray conditioning and detector calibration – are required for the X-ray source and
detector to perform at their highest levels.
Note: Be sure nothing is in the cabinet prior to detector calibration. If an item is left inside the
cabinet, detector calibration will not begin.
This is a reproduction of the X-ray System Maintenance screen. The components of this screen
are described below.
Begin Conditioning Button
Conditioning maintains the X-ray tube and prepares it for scanning. You may not begin a scan
without the X-rays being conditioned; you will be notified on the main screen if conditioning is
necessary before you begin scanning. It is important to note that conditioning can be time
consuming. IMTEC highly recommends setting automated conditioning to take place several
hours before you intend to begin scanning.
To begin manual conditioning, tap the X-ray System Maintenance button on the main screen, and
then tap the Begin Conditioning button.
As conditioning takes place a progress bar appears, the X-ray's mean value is displayed, and the
indicator icons change accordingly.