Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A 3 Ranger™ Blood/Fluid Warming Unit, Model 245 Service Manual For I.V. Use 3 Ranger™ Blood/Fluid Warming Unit, Model 247 Service Manual Not for I.V. Use 1 Model 245 • Blood/Fluid Warming Unit • For I.V. Use 1 Model 247 • Irrigation Fluid Warming Unit • Not for I.V.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A Table of Contents Technical Service and Order Placement.................................................................................................3 In-warranty Repair and Exchange................................................................................................3 When You Call for Technical Support................................................
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A English 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Technical Service and Order Placement USA TEL: FAX: +1-952-947-1200 +1-800-733-7775 +1-952-947-1400 Outside of the USA Contact your local 3M Patient Warming representative.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. English www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Introduction The Ranger warming system includes a warming unit and a disposable warming set. The Ranger warming unit is designed to be mounted to an I.V. pole. A handle located on the top of the unit makes transport easy. When mounted to the I.V.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A English 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Voltage, Alternating Current (AC) An equipotentiality plug (grounded) conductor other than a protective earth conductor or a neutral conductor, providing a direct connection between electrical equipment and the potential equalization busbar of the electrical installation.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. English www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Explanation of Signal Word Consequences WARNING: Indicates a hazardous situation which, if not avoided, could result in death or serious injury. CAUTION: Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A English 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems NOTICE: 1. Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional. 2. To avoid device damage: • Do not clean the warming unit with solvents. Damage to the case, label, and internal components may result.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. English www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Product Description The Ranger blood/fluid, and irrigation fluid warming systems consist of a warming unit and a sterile disposable fluid warming set. Slot for warming cassette Alert indicator light Model 245 I.V.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A English 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Ranger Models 245 and 247 18. 1. 3. 17. 7. 5. 11. 8. 6. 19. 13. 9. 4. 10. 20. 12. 14. 16. 21. 15. 2. 1. Upper Enclosure 8. Thermistor Sensor 15. Microswitch 2. Lower Enclosure 9. Pole Clamp 16. Alarm Board 3. Upper Heater 10. Pole Clamp Knob 17. Controller 4.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. English www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Maintenance and Storage Cleaning the Ranger Warming Unit Clean the Ranger warming unit on an as-needed basis. CAUTION 1. To reduce the risks associated with cross-contamination: • The cleaning tool provides only superficial cleaning, it does not disinfect or sterilize the interior of the unit.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems English method 1. Unplug the warming unit. 2. Unfold the cleaning tool. Wet the foam pads with a nonabrasive solution, such as AlconoxTM brand detergent. 3. Insert the tool from the back of the unit and pull the tool all the way out from the front. 4. Rinse the tool with water and repeat 3 times.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. English 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Model 245 Blood/Fluid Warming System Preventative Maintenance Checklist caution Temperature adjustments cannot be made to the Ranger warming unit. If the test shows the temperature is out of specification, please call 3M Patient Warming technical support at 1-800-733-7775. Tools and equipment • Thermocouple: 0.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems 1. Use aluminum tape to affix a thermocouple to the top surface of the Ranger warming as shown in Figure 1. Slide the cassette into the slot of the Ranger warming unit. 3. Using room temperature saline/water, prime the cassette per instructions included with it.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. English www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Testing the Over-temperature Alert Points (43°C [±1] and 46°C [+3, -2]) Model 245 The over-temperature alert test causes the Ranger warming unit to go into an over-temperature test mode by manually overriding the primary control system and engaging the heaters.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems English Ranger Blood/Fluid Warming Unit, Model 245 Maintenance Log Model 245 The operating temperature and alerts should be tested at least every 6 months or according to institutional protocol.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. English www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Ranger Irrigation Fluid Warming System, Model 247 Preventative Maintenance Checklist Caution Temperature adjustments cannot be made to the Ranger irrigation fluid warming unit.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A English 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Testing the Over-temperature Alert Points (48°C [±2] and 50°C [±2]) The over-temperature alert test causes the Ranger irrigation warming unit to go into an over-temperature test mode by manually overriding the primary control system and engaging the heaters.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. English www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Ranger Irrigation Fluid Warming, Model 247 Maintenance Log The operating temperature and alerts should be tested at least every 6 months or according to institutional protocol.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems English Replacing the Power Fuse The power fuses are located in the power entry module. tools and equipment Small slotted screwdriver method 1. Disconnect the warming unit from the power source. 2. Remove the power cord. 3. Locate the fuse carrier in the power entry module. 4.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. English www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Troubleshooting CONDITION Nothing illuminates on the warming unit panel. CAUSE SOLUTION • Unit is not turned on, plugged in, or power cord is not plugged into an appropriate outlet. • Turn unit on. Make sure the power cord is plugged into the power entry module of the warming unit.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems CONDITION • Unit alerts soon after plugging it in (unit does not have to be turned on for this condition to occur). CAUSE Over-temperature screw on bottom of unit is loose or missing.
Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. English www.rangerfluidwarming.com reorder #202524A 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems Specifications physical characteristics warming unit 4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long; wt.: 7 lb. 7 oz. (3.4 kg) certifications IEC/EN 60601-1; CAN/CSA-C22.2, No. 601.
Made in the USA by 3M Health Care. 3M is a trademark of 3M Company, used under license in Canada. RANGER, BAIR HUGGER, and the BAIR HUGGER logo are trademarks of Arizant Healthcare Inc., used under license in Canada. ©2013 Arizant Healthcare Inc. All rights reserved. 3M Deutschland GmbH, Health Care Business Carl-Schurz-Str. 1, 41453 Neuss, Germany 3M Health Care, 2510 Conway Ave., St. Paul, MN 55144 USA TEL 800-228-3957 | www.rangerfluidwarming.