5. Select Filter
This stethoscope comes equipped with three filters: bell (low frequency), diaphragm (high frequency), and
“extended” (overlap of bell and diaphragm filters). The default setting is the diaphragm filter. Instructions for
changing the default filter setting are given in the Custom Configuration section below.
To select filter: Depress and release filter button (as shown in diagram) until desired filter mode appears on the
LCD display.
Filter Button Bell Filter Icon Diaphragm Filter Icon
Extended Filter Icon
6. Adjust Sound Amplification Level
The Model 3200 sound level can be amplified in 8 increments up to 24X amplification of a non-electronic
(cardiology-level) stethoscope. Level 1 is equal to a non-electronic stethoscope. Level 9 is equal to 24X
amplification of a non-electronic stethoscope. The greater the amplification, the more bars you will see.
A default start-up amplification level can be set using the instructions provided in the Custom Configuration
section below. The default setting is marked by the box on the vertical bar shown in the volume icon in the LCD
display. The illustration shows a default amplification setting at Level 3.
Increase Amplification: Press ( + ) button until desired amplification level is achieved
Decrease Amplification: Press ( - ) button until desired amplification level is achieved