Product Diagram

25
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Not applicable
Not applicable
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the No
touch Infrared Body Thermometer
requires continued operation during
power mains interruptions, it is
recommended that the No touch
Infrared Body Thermometerbe
powered from an uninterruptible
power supply or a battery.
Power frequency
(50Hz/60Hz) magnetic
field IEC 61000-4-8
30A/m
30A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacture’s declaration – electromagnetic immunity
The No touch Infrared Body Thermometer is intended for use in the electromagnetic environment
specified below. The customer or the user ofthe No touch Infrared Body Thermometer should assure that
it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Not applicable
Not applicable
Portable and mobile RF communications equipment
should be used no closer to any part of the No touch
Infrared Body Thermometer, including cables, than the
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d= 1.2P 80 MHz to 800 MHz
d= 2.3P 800 MHz to 2.7 GHz