Product Diagram

8
- Output: DC 5 V; 1 A
- Rated input voltage shall not exceed 250 V
- Classification of protection against electric shock: Class II
- The adapter need to comply with IEC 60601-1, and provide 2 MOPP (two
means of patient protection) between primary circuit and secondary circuit.
Lithium Battery
Rechargeable Lithium Battery for Massager:
1) During the use of the therapy device, when the beep is “spit” for 3 consecutive
times, it means that the built-in lithium battery is low, you can use the adapter to
charge, put the toggle switch to “charge position, the power indicator flashes
green. When the battery is fully charged, the power indicator indicates that the
green light is always on, and it takes about 3 - 4 hours to fully charge.
2) It can be used for 5 - 7 days after full charge (30 minutes every 4 days).
3) Under normal use, the life of the rechargeable battery is > 500 charge and
discharge, it can be used for about 2 - 3 years (by 30 minutes per day).
4) When the rechargeable battery is not used for a long time, the life of the
rechargeable battery will be shortened. It is recommended to charge the battery
at least once a month.
5) The rechargeable battery can only be replaced by a service person. Internally
Lithium battery, DC 3.7 V, 1500 mAh. (Manufacture: Shenzhen EPT Battery Co.,
Ltd, Model:103048PH)
Battery for Remote Control:
1) The remote control uses the “CR2032” button battery (Lithium battery, DC 3 V,
the battery need to comply with IEC 60086-4), which can be used for about 60
days. (by 30 minutes per day).
2) When the button is not working, it indicates that the battery is low. Please
replace it with a new one.
3) When not using the product for a long time, remove the battery. Otherwise, it
may cause the battery to leak and cause the machine to malfunction.
4) Do not use the machine in an environment exceeding 40
o
C, otherwise it will
affect the performance and life of the battery.
Warning:
Do not use batteries other than the specified type. Incorrect replacement of
lithium batteries would result in an unacceptable risk.