Fluke 192B - 196B/C - 199B/C ScopeMeter Getting Started 4822 872 30701 Sep 2002 Rev.2 12/04 © 2002 Fluke Corporation, All rights reserved. Printed in the Netherlands All product names are trademarks of their respective companies.
Table of Contents Title Unpacking the Test Tool Kit ........................................................................................... Introduction .................................................................................................................... Contacting Fluke ............................................................................................................ Safety Information: Read First..................................................................................
Fluke 192B - 196B/C - 199B/C Getting Started Unpacking the Test Tool Kit The following items are included in your test tool kit: 1. 2. 3. 4. 5. 6. 7. 8. ScopeMeter Test Tool Battery Charger 10:1 Voltage Probe Set (Red) 10:1 Voltage Probe Set (Gray) Test Lead Set (Red and Black), Probe Ground Lead Getting Started Manual (this book) CD ROM with complete Users Manual Shipment box (basic version only) 6 1 2 9 7 10 e (2x) a f (2x) c (2x) d (1x) b (2x) (2x) 3 4 11 The -S versions also include: 9.
Getting Started Introduction Safety Information: Read First This Getting Started Manual provides basic information on Models 192B, 196B, 196C, 199B and 199C. Refer to the Users Manual on the accompanying CD-ROM for complete operating instructions. The Fluke Models 192B, 196B, 196C, 199B and 199C ScopeMeter test tools (hereafter referred to as “test tool”) comply with: Contacting Fluke To locate an authorized service center, visit us on the World Wide Web at: www.fluke.
Fluke 192B - 196B/C - 199B/C Getting Started Warning To avoid electrical shock or fire: • Use only the Fluke power supply, Model BC190 (Battery Charger / Power Adapter). • Before use check that the selected/indicated voltage range on the BC190 matches the local line power voltage and frequency. • For the BC190/808 universal Battery Charger/Power Adapter use only line cords that comply with the local safety regulations.
Getting Started Safety Information: Read First • Always use the test tool only in the manner specified. Max. Input Voltages Input A and B directly ............................... 300 V CAT III Input A and B via 10:1 probe ................... 1000 V CAT II 600 V CAT III METER/EXT TRIG inputs........................ 1000 V CAT II 600 V CAT III Max. Floating Voltage From any terminal to earth ground .......... 1000 V CAT II 600 V CAT III Between any terminal ..............................
Fluke 192B - 196B/C - 199B/C Getting Started Preparing for Use At delivery, the installed NiMH batteries may be empty and must be charged for 4 hours (with the test tool turned off) to reach full charge: • use only the Fluke Battery Charger/Power Adapter model BC190 • before use check that the BC190 voltage and frequency range match the local line power range • connect the battery charger to the ac outlet • connect the battery charger to the appropriate input on the test tool near Input B.
Getting Started Changing the Information Language Changing the Information Language Using the Scope During operation of the test tool messages may appear on the screen (for instance if you select a recorder function). You can select the language in which messages will be displayed. Connect the probe(s) as shown in figure 2. Display the USER key labels. Open the LANGUAGE menu to select a language. Adjusting Contrast and Brightness Display the USER key labels. Enable the arrow keys for adjustment.
Fluke 192B - 196B/C - 199B/C Getting Started Scope Readings and Waveforms Automatic scope measurement results are shown as Reading 1 (READING 1) and Reading 2 (READING 2) at the upper left and right display edge. Scope readings on (ON) or off (OFF). Glitch Detect: glitch capture on or off. Average: waveform averaging on or off. Waveform: Normal : normal waveform display. Persistence... : waveform persistence. Digital Persistence: persistence off, short, medium, long, infinite (C-versions only).
Getting Started Multimeter Input A and Input B Settings Multimeter Connect the test leads (Figure 1, item 5) to the 4-mm safety banana jack METER input. Input A (B) on or off. Input coupling AC or DC. Figure 3. Meter connections Probe Type: voltage, current or temperature probe. Attenuation: probe attenuation. Probe Cal...: calibration of 10:1 probe. Polarity: input polarity normal or inverted; variable input sensitivity. Bandwidth: bandwidth full, 10 kHz, 20 MHz. Select a measurement function.
Fluke 192B - 196B/C - 199B/C Getting Started Recorder Functions In Trend Plot: Reading 1 (2)... (Scope) or Measurement... First, choose a measurement in Scope or Meter mode. (Meter) show the measurement function menu. Show the recorder functions. Select a recorder function, then press to display the recorder key labels: VIEW ALL : see all stored samples. NORMAL : see the most recent 9 divisions. EXIT : exit the recorder mode. Start (RUN) or stop (STOP) recording. Show the recorder options.
Getting Started Zoom Zoom Making Cursor Measurements Use the zoom function to obtain a more detailed view of a waveform. Cursors allow you to make precise digital measurements on live, on recorded, and on saved waveforms. Display the ZOOM key labels. Observe that the trace is frozen (HOLD appears at the top of the screen). Display the CURSOR key labels. Select cursor measurement type: measure signal height at one point of time. Zoom in or zoom out on the waveform. Scroll.
Fluke 192B - 196B/C - 199B/C Getting Started Cursor measurements on trace A, B or M if A and B are on; M if Mathematics... is on, see ScopeReadings and Waveforms on page 6. OFF: Switch cursor measurements off. Triggering Triggering tells the test tool when to begin displaying the waveform. Automatic triggering assures a stable display of virtually any signal: Perform an auto set. AUTO appears at the top right of the screen. Input A is the default trigger source.
Getting Started Saving/Printing Screens and Set-ups instrument is set to HOLD. Saving/Printing Screens and Set-ups Press You can save screens, recordings and setups to memory, and recall them again from memory. You can print actual or recalled screens. to arm for a new trigger. Noise reject Filter: the noise reject filter reduces jitter on the screen when triggering on noisy waveforms. th NCycle: Each N trigger results in a trace update (C versions). Video on A...
Fluke 192B - 196B/C - 199B/C Getting Started User Options Display the USER key labels. Battery refresh Start a battery refresh about four times a year to keep the batteries in optimal condition. The batteries will be fully discharged and charged again. Date adjust... Open the options menu. Set the date (Year: Month: Day: ) and the date format (DD/MM/YY is day- month-year, MM/DD/YY is month-dayyear). Time Adjust... Auto Set Adjust...
Getting Started Sikkerhet (Norwegian Safety Text). Sikkerhet (Norwegian Safety Text). Se forklaring i håndboken Dobbelt isolasjon (Sikkerhetsklasse) Les nøye gjennom sikkerhetsinformasjonen nedenfor før du bruker testverktøyet. Informasjon om avhending Jord Håndboken inneholder advarsler og forsiktighetsregler der det er nødvendig å varsle om spesielle forhold. Informasjon om resirkulering Conformité Européenne “Advarsel” gjelder forhold og handlinger som kan medføre fare for brukeren.
Fluke 192B - 196B/C - 199B/C Getting Started Advarsel For å unngå elektrisk støt eller brann: • Bruk kun Fluke spenningsadapter modell BC190 (Batterilader/nettadapter). • Før bruk sjekkes at det valgte/indikerte området på BC190 er i overensstemmelse med spenning og frekvens lokalt. • Bruk bare nettledninger som tilfredsstiller lokale sikkerhetbestemmelser sammen med BC190/808 universell batterilader/nettadapter.
Getting Started Sikkerhet (Norwegian Safety Text). • Ikke benytt spenninger som varierer med mer enn 600 V fra hverandre til de isolerte inngangene når du utfører målinger i et CAT III-miljø. Ikke benytt spenninger som varierer med mer enn 1000 V fra hverandre til de isolerte inngangene når du utfører målinger i et CAT II-miljø. • Ikke benytt spenninger utover bruksområdet til instrumentet. Vær forsiktig når du bruker 1:1 testledninger fordi probespisspenningen overføres direkte til testverktøyet.
Fluke 192B - 196B/C - 199B/C Getting Started Hvis sikkerhetsfunksjonene svekkes Hvis testverktøyet brukes på annen måte enn angitt, kan beskyttelsen utstyret gir, svekkes. Før bruk må du undersøke testledningene for mekanisk skade og bytte ut skadede testledninger! Hvis det er mulighet for at sikkerheten har blitt svekket, må du slå av testverktøyet og koble den fra nettspenningen. Deretter bør kvalifisert personale undersøke saken.