User Manual

66
12-5-2. Recalling the upper and lower limit values
(Quick selection mode)
The procedure below describes an easy way to recall the upper and lower limit values to be
used for weighing. When the recalled upper and lower limit values are to be changed, refer
to "12-5-1. Storing the upper and lower limit values".
Note
Refer to "12-5-1. Storing the upper and lower limit values" for the method to store the
upper and lower limit values.
Step 1 Set the "Data memory (data)" parameter to "4".
Step 2 Press the MODE key to enter the upper/lower
limit value recalling mode. The upper limit value
last selected with its comparator number appears.
The display is as shown to the right, with all the
digits blinking.
Step 3 Press the MODE key to select the value. Each
time the MODE key is pressed, the displayed
value changes as follows:
(
........
C03 HI C03 LO C04 HI
C04 LO
........
)
Step 4 Press the PRINT key to confirm the selection. The
balance returns to the weighing mode with the
selected upper and lower limit values ready for use.
To cancel the selection, press the CAL key. The
balance returns to the weighing mode.
Note
When no operation is performed in step 4 (after a few seconds of inactivity), the
balance selects the value currently displayed and returns to the weighing mode
automatically.
Comparator
numbe
r
Upper
limit
Store
Comparison with
no. 04
Lower
limit
Target
Number
Quick selection