3 . When dav 0v is displayed, adjusting 0V is enabled.
In the right side of the sub-display, the internal correction factor for
the DAV output appears.
Use the following keys to adjust the output voltage while looking at
the digital multimeter.
1> Switches among 0 V, 10 V and –10 V.
1 Increases the output voltage value.
Also increases the correction factor.
When kept pressed, increases the correction factor by ten.
1 Decreases the output voltage value.
Also decreases the correction factor.
When kept pressed, decreases the correction factor by ten.
1 Saves the adjusted value.
ESC Cancels the operation and returns to the dav display.
Note: Before fine adjustment, the color of the main display is green.
It changes to orange when fine adjustment is performed and
the correction factor is changed. (When the correction factor
is restored to the previous value, the display color returns to
green.) Press the 1 key to save the result of fine
adjustment.
This mode directly changes the DAV output factory setting
values. The correction factor varies depending on the
indicator. Before changing the value, record the factory
setting values.
To return to
the measurement mode
dav 0v
valUe 32768
dav 0v
valUe 32496
------
valUe 32496
dav 0v
valUe 32496
dav10v
valUe 62603
dav-10
valUe 62603
ESC
dav
ESC