Instruction Manual

HV-G/HW-G Series Page 29
8. Counting Mode
8. Counting Mode
The counting mode is the function to convert the total mass value (total weight) of
articles to a count, when each article has the same mass value.
To use this function, store a unit mass in advance.
Even if the AC adapter or the batteries is removed, the unit mass is maintained in
non-volatile memory.
8.1. Storing a Unit Mass
Step 1 Press the MODE key to display the unit .
Step 2 Press the SET key to enter the mode that stores a
unit mass.
Step 3 Press the key to select the number of samples.
The greater the quantity of samples, the greater the
accuracy of the count.
5 pieces, 10 pieces, 20 pieces, 50 pieces,
100 pieces
Step 4 Place the container on the pan.
Press the TARE key.
Step 5 Place the number of samples selected at step 3.
Wait for the stability mark to be displayed. Press the
ENTER
key to store. The count is displayed.
Caution
When the sample total mass value is too small and
it is not possible to calculate a unit mass, the scale
displays
lo ut and returns to step 3. Increase the
number of samples. More than 5 digits of the total
sample mass, in the unit of kg, are required.
When the unit mass is too light to store, the scale
displays lo ut . In this case, the unit mass will not be
stored even if the number of samples is increased.
Pressing the ENTER key, after lo ut is
displayed, displays the next unit.
Step 6 Remove the samples and the container from the pan.