Instruction Manual

Page 52 HV-G/HW-G Series
15. The Function Table
15.2. Parameter List
Item Display Description
The type L scale is turned off after 5 minutes of no operation,
when zero is displayed.
f1 0 # Not used
Automatic
turning off
f1 1
Used
Selects how to change the weighing range for HV-L and HV-G
series.
f2 0 # Automatic range
Weighing
range
f2 1
Manual range
Selects the first unit at the time when the scale turns on.
f3 0 # Kg
f3 1
Lb
f3 2
Oz
Unit
f3 3
lb-oz
Selects the transmission rate of the serial interface (RS-232C/ 422/485).
f4 0 # 2400 bps
f4 1
4800 bps
Baud rate
f4 2
9600 bps
Selects the mode for the serial interface (RS-232C/ 422/485).
f5 0 # Stream mode. (Refer to "16.2. Stream Mode")
f5 1
Command mode. (Refer to "16.3. Command Mode")
f5 2
A data is output, when the PRINT key is pressed.
f5 3
Auto-print +
When the display becomes a positive stable value and not
nearly-zero, the scale outputs the data automatically. Next
output can be performed after the display becomes
nearly-zero or a negative value.
f5 4
Auto-print +/-
When the display becomes a stable value and not
nearly-zero, the scale outputs the data automatically. Next
output can be performed after the display becomes
nearly-zero.
Output
mode
f5 5
At each end of full/dribble batch function, the data is output.
Selects the comparator mode.
f6 0 #
Press the SET key to always compare the current weight value.
Press the SET key again to stop the comparison.
A
A
f6 1
The scale always compares the weight value.
A
f6 2
Press the SET key to always compare the weight value when not
nearly-zero. Press the SET key again to stop the comparison.
A
A
Comparator
f6 3
The scale always compares the weight value when not
nearly-zero.
A
"#" :factory settings. The "nearly-zero" is within ±5 digits from zero point in the unit of kg.