Instruction Manual

Page 56 HV-G/HW-G Series
16. RS-232C Serial Interface
16. RS-232C Serial Interface
Note When the RS-232C serial interface is used, be sure to set the “Address ( f11 ) “ to
“( 00 )”.
RS-232C interface has the following two modes.
Stream mode Outputs data continuously and can be used to print data.
Command mode Controls the scale using commands from a computer.
To use RS-232C interface, set the parameters of the “Baud rate ( f4 )”, “ Output
mode ( f5 )”, “ Format ( f15 )” in the function table.
Optional cables are available as follows:
AX-KO577A-200 RS-232C cable, D-sub 25 pin, 2 m
AX-KO1786-200 RS-232C cable, D-sub 9 pin, 2 m
Transmission system EIA RS-232C
Transmission Asynchronous, bi-directional, half-duplex
Data format Baud rate 2400, 4800, 9600 bps
Data 7 bits
Parity 1 bit, Even
Start bit 1 bit
Stop bit 1 bit
Code ASCII
Terminator CR LF (CR: 0Dh, LF: 0Ah)
Pin connections
Pin No. Signal name Direction Description
A
1 -
A
2 RXD Input Receive data
A
3 TXD Output Transmit data
A
4 -
A
5 SG - Signal ground
A
6 -
A
7 DSR Output Data set ready
A
8 - -
Circuits
Adaptable connector DIN 8 pin (TCP0586)