Instruction Manual

HV-G/HW-G Series Page 63
16. RS-232C Serial Interface
16.3.2. Example of Setting Parameters
Step 1 Turn off the display.
Press and hold the ZERO key and press the ON/OFF key.
The function table is displayed.
Step 2 Press the ENTER key to display f4 ("Baud rate").
Select a parameter of "Baud rate" with the key.
Step 3 Press the ENTER key to display f5 ("Output
mode").
Select "Command mode ( f5 1 )" with the key.
Step 4 Press the ENTER key to display f15 ("Format").
Select "Format 1 ( f15 0 )" with the key.
Step 5 Press the ENTER key to display end .
Step 6 Press the ENTER key to store the new parameters.
The scale returns to the normal weighing mode.
Step 7 When the computer transmits a command, the scale
replies.