SC/SE Series Wash Down Scales SC-30KAM / SC-60KAM / SC-150KAM / SC-60KAL / SC-150KAL SE-30KAM / SE-60KAM / SE-150KAM / SE-60KAL / SE-150KAL 1WMPD4002054
© 2010 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of A&D Company, Limited. The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.
CONTENTS 1. INTRODUCTION .................................................................................................. 2 2. FEATURES........................................................................................................... 2 3. UNPACKING ........................................................................................................ 3 4. DESCRIPTION OF EACH PART.......................................................................... 4 4-1. Display and symbols ................
1. INTRODUCTION This manual describes how the SC/SE Series Wash Down Scales work and how to get the most out of them in terms of performance. Please read this manual completely before using the scale. 2. FEATURES The SC/SE Series Wash Down Scales have the following features: Dust-tight and water-tight construction, complying with IP68. Water will not enter the scale even if immersed to a depth of 1 m for 24 hours.
3. UNPACKING When unpacking, check whether all of the following items are included: ●Instruction manual SC/SE Series Wash Down Scales SC-30KAM / SC-60KAM / SC-150KAM / SC-60KAL / SC-150KAL SE-30KAM / SE-60KAM / SE-150KAM / SE-60KAL / SE-150KAL ●SC-KAL / KAM series ●SE-KAL / KAM series *Pole support foot Pole support foot *Only for the SE-KAM series The cable length from the base to the display is approximately 1.5 m.
4.
●SE-KAL / KAM series Display Display pole Option slot Display rear side Battery cover Weighing pan Spirit level Leveling foot 5 Display knob
4-1. Display and symbols Display LCD STABLE indicator NET indicator Weighing units ZERO indicator ON/OFF switch Battery indicator ON/OFF HI,OK,LO indicators PRINT switch UNITS switch ZERO switch TARE switch Symbols Symbol Description STABLE ○ Turns on when the weight value is stable. NET ◄ Turns on when the NET weight is displayed. (The tare operation is in progress.) ZERO ◄ Turns on when the scale shows zero. iHIi OK iLOi Turns on when the comparator results are displayed.
4-2. Switches Switch Description ON/OFF switch Turns the power ON or OFF. When turned ON, the scale will be automatically set to zero (power-on zero). ON/OFF PRINT switch Outputs the weight value to the printer. In the setting mode, this switch is used to increase the value of the selected blinking digit by one. UNITS switch Switches the weighing unit. In the setting mode, this switch is used to shift the blinking digit to the right. ZERO switch Zeroes the scale and sets the display to zero.
5. SETTING UP 5-1. Attaching the display pole to the base SC-KAL / KAM series Lower part of the display pole 2 1. Remove the weighing pan from the base. 2. Remove the 4 screws from the lower part of the display pole. 4 3. Pull out the cable 3. Insert the display pole under the base Base while pulling out the cable into the base and align the cable opening positions of the pole and the base. (See “3” in the illustration.) Do not pull the cable forcibly.
SE-KAL / KAM series Lower part of the display pole 2 1. Remove the weighing pan from the base. 2. Remove the 4 screws from the lower part of the display pole. 4 Base 3. Insert the display pole under the base while pulling out the cable into the base and align the cable opening positions of the pole and the base. (See “3” in the illustration.) Do not pull the cable forcibly. Take care not to pinch the cable Base, seen between the pole and the base. from above 3. Pull out the cable 3.
5-2. Installing the scale 1. Select the place for installing the scale. Refer to “Precautions on installing the scale” below. 2. Adjust the level of the base, using the spirit level and leveling feet. Confirm that the bubble is in the center of the level. The SE-KAM series has an extra foot at the bottom of the display pole. Adjust this foot to touch the floor after adjusting the level of the base. Leveling foot Spirit level The size and shape depends on the scale model. 0 3.
5-3. Installing the batteries Prepare six, size D (R20P or LR20) batteries. The batteries are not included with the product. The scale can be used continuously for about 3,600 hours using alkaline batteries. When appears, replace all six batteries with new ones as soon as possible. 1. Turn the battery cover on the rear of the display counterclockwise to remove it. Battery cover 2. Press on the catch in the arrow direction to release it. Take the battery compartment out in the direction of the arrow. 3.
6. BASIC OPERATION 6-1. Turning the power ON and OFF 1. Press the ON/OFF switch to turn the power ON. All the display symbols appear and the scale waits for the weight value to become stable. After the weight value internally becomes stable, the display turns off for a moment and zero is shown with the ZERO indicator (power-on zero). If the weight value is unstable, the display shows “------”. Check if anything touches the weighing pan, or check if there is strong wind or vibration.
6-3. Basic weighing operation 1. Press the ON/OFF switch to turn the power ON. 2. Select a weighing unit using the UNITS switch. Container 3. When the display doesn’t show zero, press the ZERO switch to set the display to zero. 4. When using a tare (container), place the container on the weighing pan, and press the TARE switch to set the display to zero. 5. Place the object to be weighed on the weighing pan or in the container, and wait for the STABLE indicator to turn on and read the value. 6.
7. COUNTING MODE Determines a unit weight (the weight of one piece) from the known pieces of samples, and calculates how many pieces are on the weighing pan using the unit weight. The unit weight is maintained even if the power is turned OFF. 1. Press the UNITS Press switch to select “pcs”. (“pcs” = pieces) ZERO 2. Press and hold the UNITS switch to enter the sample unit weight storing mode. The numerical value on the left indicates the number of samples.
8. COMPARATOR Compares the weight value with the upper and lower limit values and shows the results of the comparison by HI OK LO on the display. The formula to compare is as follows: 1LO < Lower limit value ≤ OK ≤ Upper limit value < HI Select comparison conditions that suit your own application. (Refer to “Comparator mode: Cp” in “10-2. Function list”.) 0: No comparison (comparator function disabled). 1: To compare all data. 2: To compare all stable data.
Entering the upper and lower limit values 1. Press the UNITS switch to select a weighing unit, ”kg”, ”lb” or ”oz” 2. Press and hold the UNITS switch to enter the upper limit setting mode. Press Press and hold 3. Enter the upper limit value using the following switches. UNITS To shift the blinking digit to the right. PRINT To increase the value of the blinking digit by one. The minus sign can be selected as a blinking digit after the least significant digit.
9. CALIBRATION Adjusts the scale for accurate weighing. Calibrate the scale in the following cases. When the scale is first installed. When the scale has been moved. When the ambient environment has changed. For regular calibration. 9-1. Gravity acceleration correction When the scale is first used or has been moved to another location, it should be calibrated using a calibration weight. But if a calibration weight is not available, the gravity acceleration correction will compensate the scale.
9-2. Calibration using a weight Prepare a weight, preferably a weight with the same value as the weighing capacity of the scale to be calibrated. Note that the calibration weight value can be set by 10 kg/lb. 1. Turn the power ON and warm up the scale for at least half an hour. Change the function setting “poff” or place something on the weighing pan to disable the auto power-off function. Make sure that the scale is in the weighing mode (displaying “kg”, “lb”, “oz” or “pcs”).
9-3. Restoring to the factory setting values If the gravity acceleration value or calibration weight value is changed unintentionally, restore those values to the factory setting values, as follows. Make sure that the scale is in the weighing mode (displaying “kg”, “lb”, “oz” or “pcs”). 1. Press and hold the TARE switch until the gravity acceleration value appears, and release the switch. 2. Press the UNITS display “ClrC” . . switch twice and 3. Press and hold the UNITS “ClrCno” appears blinking.
10. FUNCTIONS The scale has function settings to specify the scale performance. The parameters set in the function settings are maintained even if the power is turned OFF. Parameter Item 10-1. Setting the parameters While holding down, 1. Turn the power OFF. 2. Press and hold the TARE switch and press the ON/OFF switch to turn the power ON. The software version is displayed. “∗∗∗” indicates the software version number. 3. Press the UNITS switch to display the item. press ON/OFF Press 4.
10-2.
11. OPTIONS The following options are available for the SC/SE series: OP-02 (SCE-02) OP-03 (SCE-03) USB interface RS-232C serial interface and Comparator relay output Note When an option is installed, the dust-tight and water-tight performance of the scale will be degraded. For details, refer to the instruction manual for the options. 12. MAINTENANCE 12-1. Notes on maintenance Do not disassemble the scale. Contact your local A&D dealer if the scale needs service or repair.
13. SPECIFICATIONS 13-1. Specifications Materials SC series: SE series: MODEL Weighing capacity (kg) Minimum display “d” (kg) Weighing capacity (lb) Minimum display “d” (lb) Weighing capacity (oz) Minimum display “d” (oz) Number of samples Maximum count Minimum unit weight Repeatability (Std.
13-2. External dimensions ●SC-30KAM / SC-60KAM / SC-150KAM 163 380 717 60 38.1 117 300 173 300 220 ●SC-60KAL / SC-150KAL 158 530 717 60 38.
●SE-30KAM / SE-60KAM / SE-150KAM 60 380 722 202 40 118 300 202 220 300 ●SE-60KAL / SE-150KAL 202 530 722 60 40 128 390 440 300 Unit: mm 25
GRAVITY ACCELERATION MAP Values of gravity at various locations Amsterdam Athens Auckland NZ Bangkok Birmingham Brussels Buenos Aires Calcutta Cape Town Chicago Copenhagen Cyprus Djakarta Frankfurt Glasgow Havana Helsinki Kuwait Lisbon London (Greenwich) Los Angeles Madrid 9.813 m/s2 9.807 m/s2 9.799 m/s2 9.783 m/s2 9.813 m/s2 9.811 m/s2 9.797 m/s2 9.788 m/s2 9.796 m/s2 9.803 m/s2 9.815 m/s2 9.797 m/s2 9.781 m/s2 9.810 m/s2 9.816 m/s2 9.788 m/s2 9.819 m/s2 9.793 m/s2 9.801 m/s2 9.812 m/s2 9.796 m/s2 9.
World map 27