Instruction manual

55
MODE Key Function During Measurement (fnc)
Parameter
Settings Description
0
Switches the viscosity
units.
Each time the MODE key is pressed, the viscosity unit is
switched.
SV-1A / 1H / 10A / 10H: mPas Pas, cP P
SV-100A / 100H : Pas P
Note
With the SV-1A / 1H / 10A / 10H, for the viscosity over
1000 mPas, the unit is fixed to Pas and for the viscosity
over 1000 cP, the unit is fixed to P.
1
Switches between the
temperature display
and the measurement
elapsed time display
Each time the MODE key is pressed, the display is switched
between the temperature display and the measurement
elapsed time display.
Upon a measurement start, the temperature display is
selected.
When the elapsed time reaches 100 hours, the display
returns to 0. (99.59.5900.00.00)
2
Do not display the
digits that are beyond
the decimal point
When measuring, by pressing the MODE key each time, it
does not display the digits that are beyond the decimal
point.
But with 100-999mPas, always displays the digits that are
beyond the decimal point
When using with the portable type, use this function when
you need to conceal the large variation in digits.
Data Output Mode (prt)
The condition to output the measurement data via RS-232C is set.
Parameter
Settings Description
0
Key mode
During measurement or in the data hold mode, pressing the
PRINT key outputs the current measurement values. *2
1
Auto print mode
The measurement values are output automatically when the
STOP key ends the measurement.
Pressing the PRINT key outputs the current measurement
values. *2
2
Stream mode
The measurement values are output continuously during
measurement.
When D.P. format is selected in "Output format (type)" of the
function setting, only the viscosity value is output, regardless
of the settings of "5-at", "5-td" and "5-ed".
When this mode is selected, the data hold mode using the
HOLD key is not available.
*2 When A&D standard format is selected in "Output format (type)" of the function setting,
pressing the PRINT key in the standby mode will not output the measurement data.