INSTRUCTION MANUAL Blood Pressure Monitor TM-2430-13

50
3-1-2 TM-2430 Series Error Codes
ERROR CODE POSSIBLE CAUSE CORRECTION
E00 Clock parameters have not been
set.
Enter clock parameters.
E03 An error code is displayed
without cuff inflation.
Exhaust the air from the cuff completely.
E04 Measurement is stopped due to a
low battery.
Replace with new batteries. Auto mode
is quit. Restart the Auto mode if you use
it.
E05 Inflation pressure does not reach
the target pressure.
Wrap the cuff and connect to the
monitor. If the error persists, there may
be an air leak and repair is necessary.
E06 Pressurization greater than 320
mmHg.
Do not move and try to relax during the
measurement. If you can not clear the
error, repair is necessary.
E07 Controlled stop using STOP key Do not press the STOP key when
unnecessary.
E08, E10 Pulse can not be detected. Do not move and try to relax during the
measurement.
E20 Pulse less than 35 beats or more
than 200 beats per minute.
Measure the blood pressure by other
methods.
E21 Diastolic greater than 160 mmHg
or less than 40 mmHg.
Measure the blood pressure by other
methods.
E22 Systolic greater than 280 mmHg
or less than 60 mmHg.
Measure the blood pressure by other
methods.
E23 DSD greater than 150 mmHg or
less than 10 mmHg.
(DSD = difference between
systolic blood pressure and
diastolic blood pressure)
Measure the blood pressure by other
methods.
E30, E31 Air leaks in the cuff assembly. Repair is necessary.
E32 Clock error If the error persists, repair is necessary.
E50 Pressure offset error Repair is necessary.
E52 Memory error Repair is necessary.
E53 Defective battery contact Replace batteries correctly. If the error
persists, repair is necessary.
E55, E56, E57 Exhaust error Relax and do not move during the
measurement. If this error occurs
frequently, repair is necessary.
E60 Interval setting error Enter the interval parameters correctly.
E70, 71, 72, 73 RS-232C error Re-connect the communication cable. If
the error persists, repair is necessary.
E74 Voltage decreases during
communication.
Replace batteries with new ones and
restart communication.
E75 Protocol error due to external
equipment
Re-connect the communication cable. If
the error persists, repair is necessary.