Digital Blood Pressure Monitor Model UA-705 Instruction Manual Original Manuel d’instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione 使用手冊 翻譯 1WMPD4000988D
Contents Dear Customers .............................................................................................1 Preliminary Remarks.......................................................................................1 Precautions.....................................................................................................2 Parts Identification..........................................................................................3 Symbols....................................................
The device is designed for use on adults only, not newborns or infants. Environment for use: The device is for use in the home healthcare environment. This device is designed to measure blood pressure and pulse rate of people for diagnosis. Precautions Precision components are used in the construction of this device. Extremes in temperature, humidity, direct sunlight, shock or dust should be avoided. Clean the device and cuff with a dry, soft cloth or a cloth dampened with water and a neutral detergent.
Parts Identification Display Air socket Air connector plug Start button Exhaust valve Air hose Ruber bulb Artery position mark Proper fit range Index mark Arm cuff Battery cover Battery compartment Battery LR6P, R6 or AA Part Of Display Heart mark Systolic pressure MEMORY Diastolic pressure Pressure Bar Indicator and WHO Classification Indicator Pulse rate Battery indicator Exhaust mark. (Exhaust air from the cuff) I.H.B. indicator (Irregular heartbeat indicator) Pressure mark.
Symbols Symbols Function / Meaning Recommended Action Standby and ON switch. Battery installation guide Type BF: Device, cuff and tubing are designed to provide special protection against electrical shocks. The indicator that appears while Measurement is in progress. measurement is in progress. It blinks Remain as still as possible. when the pulse is detected. Irregular Heartbeat indicator. (I.H.B.
Symbols SN Function / Meaning Recommended Action Refer to instruction manual/booklet Serial number Using The Monitor Installing / Changing The Batteries 1. Remove the battery cover. 2. Insert a new battery into the battery compartment as shown, taking care that the polarities (+) and (-) are correct. Use only R6P, LR6, AA or equivalent battery. 3. Close the battery cover. CAUTION Insert the battery as shown in the battery compartment. If not, the device will not work.
Using The Monitor Selecting The Correct Cuff Using the correct cuff size is important for an accurate reading. If the cuff is not the proper size, the reading may yield an incorrect blood pressure value. The arm size is printed on each cuff. The index and proper fit range, on the cuff, tell you if you are applying the correct cuff. (Refer to "Applying The Arm Cuff" on next page) If the index points outside of the range, contact your local dealer to purchase a replacement cuff. The arm cuff is a consumable.
Using The Monitor Applying The Arm Cuff Do not roll up shirtsleeve tightly 1. Wrap the cuff around the upper arm, about 1 to 2 cm above the inside of the elbow, as shown. Place the cuff directly against the skin, as clothing may cause a faint pulse and result in a measurement error. Cuff Artery position mark 2. Constriction of the upper arm, caused by rolling up a shirtsleeve, may prevent accurate readings. 3. Confirm that the index proper fit range.
Measurements Measurement With Storing Data 1. Place the cuff on the arm (preferably the left arm). Sit quietly during measurement. 2. Press the START button. The buzzer sounds and the average blood pressures and the WHO classification and last data number are displayed briefly. When the " 0 " is flashing, the device is ready for mark is displayed, the cuff measurement. If the has some air trapped in it. Press the exhaust valve mark is displayed. button until the 3.
If a subsequent measurement is required, turn off the power and turn it on again. When "0 " is displayed, the device is ready for another measurement. Note: Allow at least three minutes between measurements on the same person. 9. Automatic power-off function If the device is left on after a measurement, it will turn off automatically after about one minute. It may be turned off at any time by pressing the START button. Notes For Proper Measurement Sit down in a comfortable position.
WHO Classification Indicator Each of the six segments of the bar indicator corresponds to the WHO blood pressure classification described on the next page. WHO Classification Indicator Severe hypertension Moderate hypertension Mild hypertension High normal Normal Optimal Example: Moderate hypertension : The indicator displays a segment, based on the current data, corresponding to the WHO classification.
What Is Hypertension And How Is It Controlled? Hypertension, an abnormally high arterial blood pressure, if left unattended, can cause many health problems including stroke and heart attack. Hypertension can be controlled by altering one's lifestyle, avoiding stress, and with medication under a doctor’s supervision.
Troubleshooting Problem Nothing appears in the display, even when the power is turned on. Possible Reason Recommended Action Battery is drained. Replace all battery with a new one. Battery polarities are Reinstall the battery with its negative and not in the correct positive end matching those indicated in position. the battery compartment. The cuff is not fastened Fasten the cuff correctly. properly. You moved your arm Make sure you remain very still and quiet or body during the during the measurement.
Technical Data Type UA-705 Measurement method Oscillometric measurement Measurement range Pressure: Pulse: 20 - 280 mmHg 40 - 200 beats / minute Measurement accuracy Pressure: Pulse: ±3 mmHg ±5% Power supply 1 x 1.5V batteries (R6P, LR6 or AA) Number of measurements Approximately 4000 measurements, when AA Alkaline batteries are used, with pressure value of 180 mmHg at room temperature of 23 °C.
Asahi 1-243, Kitamoto-shi, Saitama 364-8585, JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119 A&D INSTRUMENTS LIMITED Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485 A&D ENGINEERING, INC. 1756 Automation Parkway, San Jose, California 95131, U.S.A.