Instruction manual

English 6
Symbols
Symbols Function / Meaning Recommended Action
Appears during measurement when
the cuff is attached loosely
The reading may yield an
incorrect value. Apply the
cuff correctly, and take
another measurement.
Detected rate of IHB in memory
User
Previous measurements stored in memory.
Average data
Full Battery
The battery power indicator during
measurement.
Low Battery
The battery is low when it blinks.
Replace all batteries with new
ones when the mark blinks.
Unstable blood pressure due to
movement during measurement.
Take another measurement.
Remain very still during
measurement.
The systolic and diastolic values are
within 10 mmHg of each other.
The pressure value did not increase
during the inflation.
The cuff is not applied correctly.
PUL. DISPLAY ERROR
The pulse is not detected correctly.
Apply the cuff correctly,
and take another
measurement.
Blood pressure monitor internal error
Remove the batteries and
press the
START button,
and then install the batteries
again. If the error still
appears, contact the dealer.
AM
Data taken between 4:00 and 9:59
PM
Data taken between 18:00 and 1:59
%IHB =
x 100 [%]
Number of detected
IHBs in memory
Total number